Factors Related to Proximal Junctional Kyphosis and Device Failure in Patients with Early-Onset Scoliosis Treated with a Traditional Dual Growing Rod: A Single Institution Study.

Mohammadreza Chehrassan, Farshad Nikouei, Mohammadreza Shakeri, Javad Moeini, Behnam Jafari, Ebrahim Ameri Mahabadi, Hasan Ghandhari

Open access · diamondAbstract read
In one paragraph

Article in Asian spine journal, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
1.7field-weighted citation impact, top 18% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed, 3 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 1 institution in 1 country.

Mohammadreza ChehrassanBone and Joint Reconstruction Research Center, Shafa Orthopaedic Hospital, Iran University of Medical Sciences, Tehran, Iran.
Farshad NikoueiBone and Joint Reconstruction Research Center, Shafa Orthopaedic Hospital, Iran University of Medical Sciences, Tehran, Iran.
Mohammadreza ShakeriBone and Joint Reconstruction Research Center, Shafa Orthopaedic Hospital, Iran University of Medical Sciences, Tehran, Iran.
Javad MoeiniBone and Joint Reconstruction Research Center, Shafa Orthopaedic Hospital, Iran University of Medical Sciences, Tehran, Iran.
Behnam JafariBone and Joint Reconstruction Research Center, Shafa Orthopaedic Hospital, Iran University of Medical Sciences, Tehran, Iran.
Ebrahim Ameri MahabadiBone and Joint Reconstruction Research Center, Shafa Orthopaedic Hospital, Iran University of Medical Sciences, Tehran, Iran.
Hasan GhandhariBone and Joint Reconstruction Research Center, Shafa Orthopaedic Hospital, Iran University of Medical Sciences, Tehran, Iran.
Iran University of Medical Sciences · IR

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

STUDY

designObservational study. PURPOSE: Investigation of factors related to proximal junctional kyphosis (PJK) and device failure in patients with early-onset scoliosis. OVERVIEW OF LITERATURE: The use of growth-friendly devices, such as traditional dual growing rod (TDGR) for the treatment of earlyonset scoliosis (EOS), may be associated with important complications, including PJK and device failure.

methodsThirty-five patients with EOS and treated with TDGR from 2014 to 2021 with a minimum follow-up of 2 years were retrospectively evaluated. Potential risk factors, including demographic factors, disease etiology, radiological measurements, and surgical characteristics, were assessed.

resultsPJK was observed in 19 patients (54.3%), and seven patients (20%) had device failure. PJK was significantly associated with global final kyphosis change (p=0.012). No significant correlation was found between the rod angle contour, type of implant, connector design, and the risk of PJK or device failure.

conclusionsTreatment of EOS with TDGR is associated with high rates of complications, particularly PJK and device failure. The device type may not correlate with the risk of PJK and device failure. The progression of thoracic kyphosis during multiple distractions is an important risk factor for PJK.

Indexed as

Early-onset scoliosisEquipment failure CopyrightProximal junctional kyphosisTraditional dual growing rod

Identifiers

PMID38454749
PMCPMC11065516
OpenAlexW4392591126

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.