Evidence map›Paper›PMID 38404809›Full record

ArticleHeliyon2024

Unilateral external fixator and its biomechanical effects in treating different types of femoral fracture: A finite element study with experimental validated model.

Aishah Umairah Abd Aziz, Muhammad Imam Ammarullah, Bing Wui Ng, Hong-Seng Gan, Mohammed Rafiq Abdul Kadir, Muhammad Hanif Ramlee

Open access · goldAbstract read
In one paragraph

Article in Heliyon, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed
10.4field-weighted citation impact, top 1% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

8 citing papers in PubMed, 27 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 5 institutions in 3 countries.

Aishah Umairah Abd AzizBone Biomechanics Laboratory (BBL), Department of Biomedical Engineering and Health Sciences, Faculty of Electrical Engineering, Universiti Teknologi Malaysia, Johor Bahru, 81310, Johor, Malaysia.
Muhammad Imam AmmarullahDepartment of Mechanics and Aerospace Engineering, College of Engineering, Southern University of Science and Technology, Shenzhen, 518055, Guangdong, China.
Bing Wui NgDepartment of Orthopaedics and Traumatology, Hospital Universiti Kebangsaan Malaysia (HUKM), Cheras, 56000, Federal Territory of Kuala Lumpur, Malaysia.
Hong-Seng GanSchool of AI and Advanced Computing, XJTLU Entrepreneur College (Taicang), Xi'an Jiaotong-Liverpool University, Suzhou, 215400, Jiangsu, China.
Mohammed Rafiq Abdul KadirDepartment of Biomedical Engineering, Faculty of Engineering, Universiti Malaya, Kuala Lumpur, 50603, Federal Territory of Kuala Lumpur, Malaysia.
Muhammad Hanif RamleeBone Biomechanics Laboratory (BBL), Department of Biomedical Engineering and Health Sciences, Faculty of Electrical Engineering, Universiti Teknologi Malaysia, Johor Bahru, 81310, Johor, Malaysia.
University of Technology Malaysia · MYNational University of Malaysia · MYSouthern University of Science and Technology · CNUniversity of Malaya · MYXi’an Jiaotong-Liverpool University · CN

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Previous works had successfully demonstrated the clinical effectiveness of unilateral external fixator in treating various types of fracture, ranging from the simple type, such as oblique and transverse fractures, to complex fractures. However, literature that investigated its biomechanical analyses to further justify its efficacy is limited. Therefore, this paper aimed to analyse the stability of unilateral external fixator for treating different types of fracture, including the simple oblique, AO32C3 comminuted, and 20 mm gap transverse fracture. These fractures were reconstructed at the distal diaphysis of the femoral bone and computationally analysed through the finite element method under the stance phase condition. Findings showed a decrease in the fixation stiffness in large gap fracture (645.2 Nmm-1 for oblique and comminuted, while 23.4 Nmm-1 for the gap fracture), which resulted in higher displacement, IFM and stress distribution at the pin bone interface. These unfavourable conditions could consequently increase the risk of delayed union, pin loosening and infection, as well as implant failure. Nevertheless, the stress observed on the fracture surfaces was relatively low and in controlled amount, indicating that bone unity is still allowable in all models. Briefly, the unilateral fixation may provide desirable results in smaller fracture gap, but its usage in larger gap fracture might be alarming. These findings could serve as a guide and insight for surgeons and researchers, especially on the biomechanical stability of fixation in different fracture types and how will it affect bone unity.

Indexed as

External fixatorFemurFinite element analysisFracture healingPin tract infection

Identifiers

PMID38404809
PMCPMC10884926
OpenAlexW4391896007

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.