Simplified Method for Removing Direct Oral Anticoagulant Interference in Mechanical Coagulation Test Systems-A Proof of Concept.

Thomas Exner, Manita Dangol, Emmanuel J Favaloro

Open access · goldAbstract read
In one paragraph

Article in Journal of clinical medicine, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
2.4field-weighted citation impact, top 11% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed, 6 citations in OpenAlex.

  1. Review
  2. Article
  3. Article
  4. Innovative Diagnostic Solutions in Hemostasis.Diagnostics (Basel, Switzerland) · 2024
    Review
  5. Review
  6. Testing for the lupus anticoagulant: the good, the bad, and the ugly.Research and practice in thrombosis and haemostasis · 2024
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors at 1 institution in 1 country.

Thomas ExnerHaematex Research Pty Ltd., Sydney, NSW 2077, Australia.
Manita DangolHaematex Research Pty Ltd., Sydney, NSW 2077, Australia.
Emmanuel J FavaloroHaematology Department, Institute of Clinical Pathology and Medical Research (ICPMR), Sydney Centres for Thrombosis and Haemostasis, Westmead Hospital, Sydney, NSW 2145, Australia.ORCID 0000-0002-2103-1661
The University of Sydney · AU

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundDirect oral anticoagulants (DOACs) cause unwanted interference in various hemostasis assays, including lupus anticoagulant (LA) testing, where false positive and false negative identification may occur. DOAC Stop (DS) is an activated charcoal (AC) product used to specifically and effectively adsorb DOACs from test plasma. This process normally requires plasma treatment, centrifugation and plasma separation prior to tests, but inexperienced operators may also inadvertently transfer residual AC particles, thereby potentially adversely affecting clot detection.

methodsWe hypothesized that residual DS might not be problematic for mechanical clot detection. We therefore investigated the potential impact of DS and a new DS liquid (DS-L) product on clotting tests using a mechanical clot detection system. Varying concentrations of DS were added to normal and abnormal plasmas with and without DOAC presence. Clotting tests including PT, APTT and dRVVT were performed directly in the analyzer without plasma/DS centrifugation.

resultsDS up to double the recommended treatment level had only minor effects on all test results, despite completely obscuring visibility in the plasma/reagent mix. This confirms that the centrifugation step may be able to be omitted when using mechanical detection systems.

conclusionsShould DS carryover into treated plasmas occur, this should not cause issues with testing performed on mechanical clot-sensing devices. Moreover, we hypothesize that DS can be used directly in these systems, without the need for centrifugation, thereby simplifying its many potential applications.

Indexed as

activated charcoalclotting testsDOACsDOAC Stopmechanical clot detection

Identifiers

PMID38398355
PMCPMC10889288
OpenAlexW4391747489

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.