Validity of BiliDx as a point-of-care bilirubin measurement device to diagnose and monitor neonatal jaundice at Muhimbili National Hospital, an observational study.
Observational study in BMC pediatrics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed, 2 citations in OpenAlex.
- Developing paediatric subspecialty training in resource-limited settings; the first neonatology subspeciality program in Tanzania.BMC medical education · 2026Article
- Skin residual bilirubin volume: a physiologically informed framework for transcutaneous bilirubin interpretation in neonates.Frontiers in pediatrics · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundNeonatal jaundice is a condition caused by elevated levels of bilirubin in the bloodstream. Laboratory determination of serum bilirubin concentration by total serum bilirubin (TSB) test is still considered as gold standard for clinical guidance and practice. In developed countries, diagnosis of neonatal jaundice is shifting towards point-of-care medical devices. BiliDx is a device developed to allow a fast, blood-based determination of bilirubin levels at the point of care. This study aimed to determine the accuracy of the BiliDx device relative to a standard laboratory total serum bilirubin to diagnose and monitor jaundice among neonates admitted at Muhimbili National Hospital (MNH). MATERIAL AND METHODOLOGY: This was a prospective hospital-based observational study conducted at the Neonatal Ward - MNH, Dar-es-Salaam, Tanzania from November 2022 to January 2023. A total of 180 neonates admitted at the neonatal ward with jaundice and whose parents consented were enrolled in the study. Blood samples were collected; 2 ml of venous blood into the vacutainer bottle for standard laboratory measurement of total serum bilirubin (TSB) and 25µL blood collected into a transfer pipette tube and applied to BiliDx. STATA version 15.1 was used for data analysis.
resultsOut of 180 neonates, 39.4% (71/180) had birth weight between 1500 - 2499.9 g, approximately 2/3rd (120/180) were preterm, 92/180 (51.1%) were males and 100/180 (55.6%) were undergoing phototherapy treatment the moment sample taken. The mean bilirubin concentration was 92 mmol/l for BiliDx and 118 mmol/l for standard laboratory TSB. The minimum and maximum values obtained with BiliDx were, 3.4 and 427.5 mmol/l respectively, compared with 10.7 and 382.1 mmol/l using standard laboratory TSB. A linear relationship and correlation coefficient of 0.8408 (p = 0.000) between BiliDx and standard laboratory TSB was found. The regression analysis showed the presence of constant error [coefficient of BiliDx/slope = 0.91, 95% CI (0.82-0.99), p = 0.000] and random error exclusively [coefficient of constant/y-intercept = 48.52, 95%CI (37.70-59.34), p = 0.000]. The Bland-Altman plot showed an acceptable mean difference of 39.1mmol/l, limits of agreement of -48.3mmol/l to 126.4mmol/l, and 179 points (179/180 = 99.4%) lying inside the limits of agreement.
conclusionThe results support the use of BiliDx for rapid and accurate testing of elevated levels of bilirubin in the bloodstream among neonates since 99.4% of the differences between BiliDx and standard laboratory TSB lie between the lines of agreement.