Assessing quality assurance of multi-leaf collimator using the structural similarity index.

Hong Zhang, Baoshe Zhang, Giovanni Lasio, Shifeng Chen, Joubin Nasehi Tehrani

Open access · goldAbstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
2.4field-weighted citation impact, top 14% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 4 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 2 institutions in 1 country.

Hong ZhangDepartments of Radiation Oncology, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
Baoshe ZhangDepartments of Radiation Oncology, Medical School, University of Maryland, Baltimore, Maryland, USA.
Giovanni LasioDepartments of Radiation Oncology, Medical School, University of Maryland, Baltimore, Maryland, USA.
Shifeng ChenDepartments of Radiation Oncology, Medical School, University of Maryland, Baltimore, Maryland, USA.
Joubin Nasehi TehraniDepartments of Radiation Oncology, Medical School, University of Maryland, Baltimore, Maryland, USA.
University of Maryland Medical Center · USVanderbilt University Medical Center · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThis study aims to evaluate the viability of utilizing the Structural Similarity Index (SSI*) as an innovative imaging metric for quality assurance (QA) of the multi-leaf collimator (MLC). Additionally, we compared the results obtained through SSI* with those derived from a conventional Gamma index test for three types of Varian machines (Trilogy, Truebeam, and Edge) over a 12-week period of MLC QA in our clinic.

methodTo assess sensitivity to MLC positioning errors, we designed a 1 cm slit on the reference MLC, subsequently shifted by 0.5-5 mm on the target MLC. For evaluating sensitivity to output error, we irradiated five 25 cm × 25 cm open fields on the portal image with varying Monitor Units (MUs) of 96-100. We compared SSI* and Gamma index tests using three linear accelerator (LINAC) machines: Varian Trilogy, Truebeam, and Edge, with MLC leaf widths of 1, 0.5, and 0.25 mm. Weekly QA included VMAT and static field modes, with Picket fence test images acquired. Mechanical uncertainties related to the LINAC head, electronic portal imaging device (EPID), and MLC during gantry rotation and leaf motion were monitored.

resultsThe Gamma index test started detecting the MLC shift at a threshold of 4 mm, whereas the SSI* metric showed sensitivity to shifts as small as 2 mm. Moreover, the Gamma index test identified dose changes at 95MUs, indicating a 5% dose difference based on the distance to agreement (DTA)/dose difference (DD) criteria of 1 mm/3%. In contrast, the SSI* metric alerted to dose differences starting from 97MUs, corresponding to a 3% dose difference. The Gamma index test passed all measurements conducted on each machine. However, the SSI* metric rejected all measurements from the Edge and Trilogy machines and two from the Truebeam.

conclusionsOur findings demonstrate that the SSI* exhibits greater sensitivity than the Gamma index test in detecting MLC positioning errors and dose changes between static and VMAT modes. The SSI* metric outperformed the Gamma index test regarding sensitivity across these parameters.

Indexed as

Radiotherapy, Intensity-ModulatedElectrical Equipment and SuppliesHumansParticle AcceleratorsPhantoms, ImagingRadiotherapy Planning, Computer-AssistedRotationGamma index testMLCQASSI*Structural Similarity IndexVMAT