ArticleCancers2024
Dose Calculation Accuracy of Beam Models in RadCalc for a 1.5 T MR-Linac.
Article in Cancers, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Evaluation of ThinkQA (v2.0.1.11) as an online secondary dose check for MR guided radiation therapy with the Elekta Unity MR-Linac.Physical and engineering sciences in medicine · 2026Article
- Beam model development and clinical experience with RadCalc for treatment plan quality assurance in online adaptive workflow with an MR-linac.Journal of applied clinical medical physics · 2025Article
- Feasibility of Monte Carlo-based patient-specific quality assurance in 1.5 Tesla magnetic resonance-guided online adaptive radiotherapy: a multi-institutional study.Physics and imaging in radiation oncology · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
Abstract
The purpose of this study is to evaluate RadCalc, an independent dose verification software, for patient-specific quality assurance (PSQA) in online adaptive planning with a magnetic resonance linear accelerator (MR-linac) of a 1.5 T. Version 7.1.4 of RadCalc to introduce the capability to establish a beam model that incorporates MR field characteristics. A total of six models were established, with one using manufacturer-provided data and the others differing in percentage depth dose (PDD) data sources. Overall, two models utilized PDD data from the treatment planning system (TPS), and three used commissioned PDD data from gantry angles of 0° and 270°. Simple tests on a virtual water phantom assessed dose-calculation accuracy, revealing percentage differences ranging from -0.5% to -20.6%. Excluding models with significant differences, clinical tests on 575 adaptive plans (prostate, liver, and breast) showed percentage differences of -0.51%, 1.12%, and 4.10%, respectively. The doses calculated using RadCalc demonstrated similar trends to those of the PSQA-based measurements. The newly released version of RadCalc enables beam modeling that considers the characteristics of the 1.5 T magnetic field. The accuracy of the software in calculating doses at 1.5 T magnetic fields has been verified, thereby making it a reliable and effective tool for PSQA in adaptive plans.