Evidence map›Paper›PMID 38319775›Full record

ArticleIEEE transactions on biomedical circuits and systems2024

A Low-Power Impedance-to-Frequency Converter for Frequency-Multiplexed Wearable Sensors.

Weilun Li, Junyi Zhao, Yong Wang, Chuan Wang, Shantanu Chakrabartty

Abstract read
In one paragraph

Article in IEEE transactions on biomedical circuits and systems, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Weilun Li
Junyi Zhao
Yong Wang
Chuan Wang
Shantanu Chakrabartty

Funding

A self-capacitance driven wearable electromyometrial imaging system for maternal and fetal monitoring during pregnancy and laborR01HD105905 · NICHD · WASHINGTON UNIVERSITY · PI Chuan Wang · 2022 to 2026
$1.7M
Gates Foundation INV-037302NICHD NIH HHS R01 HD105905
6 · The paper itself

Abstract

We propose a low-power impedance-to-frequency (I-to-F) converter for wearable transducers that change both its resistance and capacitance in response to mechanical deformation or changes in ambient pressure. At the core of the proposed I-to-F converter is a fixed-point circuit comprising of a voltage-controlled relaxation oscillator and a proportional-to-temperature (PTAT) current reference that locks the oscillation frequency according to the impedance of the transducer. Using both analytical and measurement results we show that the operation of the proposed I-to-F converter is well matched to a specific class of sponge mechanical transducer where the system can achieve higher sensitivity when compared to a simple resistance measurement techniques. Furthermore, the oscillation frequency of the converter can be programmed to ensure that multiple transducer and I-to-F converters can communicate simultaneously over a shared channel (physical wire or virtual wireless channel) using frequency-division multiplexing. Measured results from proof-of-concept prototypes show an impedance sensitivity of 19.66 Hz/ Ω at 1.1 kΩ load impedance magnitude and a current consumption of 128 μ A. As a demonstration we show the application of the I-to-F converter for human gesture recognition and for radial pulse sensing.

Indexed as

Electric ImpedanceEquipment DesignWearable Electronic DevicesHumansTransducersWireless Technology

Identifiers

PMID38319775
PMCPMC11706616

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.