ArticleJournal of applied clinical medical physics2024
Commissioning and initial validation of Eclipse eMC algorithm for the electron FLASH research extension (FLEX) system for pre-clinical studies.
Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed, 4 citations in OpenAlex.
- On the acceptance, commissioning, and quality assurance of electron FLASH units.Medical physics · 2025Article
- Characterization of a commercial plastic scintillator for electron FLASH dosimetry.Journal of applied clinical medical physics · 2024Article
- Commissioning and initial validation of Eclipse eMC algorithm for the electron FLASH research extension (FLEX) system for pre-clinical studies.Journal of applied clinical medical physics · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 1 institution in 1 country.
Funding
Abstract
purposeTo investigate the feasibility of commissioning the 16 MeV electron FLASH Extension (FLEX) in the commercial treatment planning system (TPS) for biomedical research with cell and mouse models, and in silico treatment planning studies.
methodsTo commission the FLEX system with the electron Monte Carlo (eMC) algorithm in the commercial TPS, radiochromic film was used to measure the vendor-recommended beam data. Once the beam model was generated for the eMC algorithm, supplemental measurements were collected for validation purposes and compared against the TPS-calculated results. Additionally, the newly commissioned 16 MeV FLASH beam was compared to the corresponding 16 MeV conventional electron beam.
resultsThe eMC algorithm effectively modeled the FLEX system. The eMC-calculated PDDs and profiles for the 16 MeV electron FLASH beam agreed with measured values within 1%, on average, for 6 × 6 cm
conclusionsThe FLEX system was successfully commissioned in the commercial TPS using the eMC algorithm, which accurately modeled the forward-peaked nature of the FLASH beam. A commissioned TPS for FLASH will be useful for pre-clinical cell and animal studies, as well as in silico FLASH treatment planning studies for future clinical implementation.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials