Evidence map›Paper›PMID 38303594›Full record

ArticleClinics in shoulder and elbow2024

Discovery Elbow System arthroplasty polyethylene bearing exchange: outcomes and experience.

Daniel L J Morris, Katherine Walstow, Lisa Pitt, Marie Morgan, Amol A Tambe, D I Clark, Timothy Cresswell, Marius P Espag

Open access · diamondAbstract read
In one paragraph

Article in Clinics in shoulder and elbow, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
0.2field-weighted citation impact, top 45% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed, 1 citations in OpenAlex.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 1 institution in 1 country.

Daniel L J MorrisDepartment of Trauma and Orthopaedics, University Hospitals of Derby and Burton NHS Foundation Trust, Derby, UK.
Katherine WalstowDepartment of Trauma and Orthopaedics, University Hospitals of Derby and Burton NHS Foundation Trust, Derby, UK.
Lisa PittDepartment of Trauma and Orthopaedics, University Hospitals of Derby and Burton NHS Foundation Trust, Derby, UK.
Marie MorganDepartment of Trauma and Orthopaedics, University Hospitals of Derby and Burton NHS Foundation Trust, Derby, UK.
Amol A TambeDepartment of Trauma and Orthopaedics, University Hospitals of Derby and Burton NHS Foundation Trust, Derby, UK.
D I ClarkDepartment of Trauma and Orthopaedics, University Hospitals of Derby and Burton NHS Foundation Trust, Derby, UK.
Timothy CresswellDepartment of Trauma and Orthopaedics, University Hospitals of Derby and Burton NHS Foundation Trust, Derby, UK.
Marius P EspagDepartment of Trauma and Orthopaedics, University Hospitals of Derby and Burton NHS Foundation Trust, Derby, UK.
University Hospitals of Derby and Burton NHS Foundation Trust · GB

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe Discovery Elbow System (DES) utilizes a polyethylene bearing within the ulnar component. An exchange bearing requires preoperative freezing and implantation within 2 minutes of freezer removal to allow insertion. We report our outcomes and experience using this technique.

methodsThis was an analysis of a two-surgeon consecutive series of DES bearing exchange. Inclusion criteria included patients in which exchange was attempted with a minimum 1-year follow-up. Clinical and radiographic review was performed 1, 2, 3, 5, 8 and 10 years postoperative. Outcome measures included range of movement, Oxford Elbow Score (OES), Mayo Elbow Performance Score (MEPS), complications and requirement for revision surgery.

resultsEleven DESs in 10 patients were included. Indications were bearing wear encountered during humeral component revision (n=5); bearing failure (n=4); and infection treated with debridement, antibiotics and implant retention (DAIR; n=2). Bearing exchange was conducted on the first attempt in 10 cases. One case required a second attempt. One patient developed infection postoperatively managed with two-stage revision. Mean follow-up of the bearing exchange DES was 3 years. No further surgery was required, with no infection recurrence in DAIR cases. Mean elbow flexion-extension and pronosupination arcs were 107° (±22°) and 140° (±26°). Mean OES was 36/48 (±12) and MEPS was 83/100 (±19).

conclusionsOur results support the use of DES bearing exchange in cases of bearing wear with well-fixed stems or acute infection. This series provides surgeons managing DES arthroplasty with management principles, successful and reproducible surgical techniques and expected clinical outcomes in performing DES polyethylene bearing exchange. Level of evidence: IV.

Indexed as

ArthroplastyElbowElbow jointElbow prosthesisPolyethyleneReplacement

Identifiers

PMID38303594
PMCPMC10938009
OpenAlexW4391296893

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.