An LC-MS-based designated comparison method with similar performance to the Lp(a) reference measurement procedure to guide molar Lp(a) standardization.

Nina M Diederiks, L Renee Ruhaak, Fred P H T M Romijn, Mervin M Pieterse, Nico P M Smit, Christa M Cobbaert

Open access · goldAbstract read
In one paragraph

Article in Clinical proteomics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
3.4field-weighted citation impact, top 8% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 6 citations in OpenAlex.

  1. Trial
  2. Review
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 1 institution in 1 country.

Nina M Diederiks *Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Postzone E2-P, Albinusdreef 2, 2333 ZA, Leiden, The Netherlands.ORCID http://orcid.org/0000-0001-9138-1969
L Renee Ruhaak *Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Postzone E2-P, Albinusdreef 2, 2333 ZA, Leiden, The Netherlands.ORCID http://orcid.org/0000-0003-3737-3807
Fred P H T M RomijnDepartment of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Postzone E2-P, Albinusdreef 2, 2333 ZA, Leiden, The Netherlands.
Mervin M PieterseDepartment of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Postzone E2-P, Albinusdreef 2, 2333 ZA, Leiden, The Netherlands.
Nico P M SmitDepartment of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Postzone E2-P, Albinusdreef 2, 2333 ZA, Leiden, The Netherlands.ORCID http://orcid.org/0000-0002-8166-5332
Christa M CobbaertDepartment of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Postzone E2-P, Albinusdreef 2, 2333 ZA, Leiden, The Netherlands. C.M.Cobbaert@lumc.nl.ORCID http://orcid.org/0000-0003-3565-1404
Leiden University Medical Center · NL

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe 2022 consensus statement of the European Atherosclerosis Society (EAS) on lipoprotein(a) (Lp(a)) recognizes the role of Lp(a) as a relevant genetically determined risk factor and recommends its measurement at least once in an individual's lifetime. It also strongly urges that Lp(a) test results are expressed as apolipoprotein (a) (apo(a)) amount of substance in molar units and no longer in confounded Lp(a) mass units (mg/dL or mg/L). Therefore, IVD manufacturers should transition to molar units. A prerequisite for this transition is the availability of an Lp(a) Reference Measurement Procedure (RMP) that allows unequivocal molecular detection and quantification of apo(a) in Lp(a). To that end an ISO 17511:2020 compliant LC-MS based and IFCC-endorsed RMP has been established that targets proteotypic peptides of apolipoprotein(a) (apo(a)) in Lp(a). The RMP is laborious and requires highly skilled operators. To guide IVD-manufacturers of immunoassay-based Lp(a) test kits in the transition from mass to molar units, a Designated Comparison Method (DCM) has been developed and evaluated.

methodsTo assess whether the DCM provides equivalent results compared to the RMP, the procedural designs were compared and the analytical performance of DCM and RMP were first evaluated in a head-to-head comparison. Subsequently, apo(a) was quantified in 153 human clinical serum samples. Both DCM and RMP were calibrated using external native calibrators that produce results traceable to SRM2B. Measurement uncertainty (MU) was checked against predefined allowable MU.

resultsThe major difference in the design of the DCM for apo(a) is the use of only one enzymatic digestion step. The analytical performance of the DCM and RMP for apo(a) is highly similar. In a direct method comparison, equivalent results were obtained with a median regression slope 0.997 of and a median bias of - 0.2 nmol/L (- 0.2%); the intermediate imprecision of the test results was within total allowable error (TEa) (CVa of 10.2% at 90 nmol/L).

conclusionsThe semi-automated, higher throughput, LC-MS-based method for Lp(a) meets the predefined analytical performance specifications and allowable MU and is hence applicable as a higher order Designated Comparison Method, which is ideally suited to guide IVD manufacturers in the transition from Lp(a) mass to molar units.