Receive coil quality assurance procedure and automated analysis for ViewRay MRIdian MR-Linac.

Evangelia Kaza, Christopher L Williams

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors.

Evangelia KazaRadiation Oncology, Brigham and Women's Hospital, Dana-Farber Cancer Institute, Harvard Medical School, Boston, Massachusetts, USA.
Christopher L WilliamsRadiation Oncology, Brigham and Women's Hospital, Dana-Farber Cancer Institute, Harvard Medical School, Boston, Massachusetts, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeRegular receiving coil quality assurance (QA) is required to ensure image quality of an MRIdian Linac system. The manufacturer provides a spherical phantom and positioning tube for single-slice signal-to-noise ratio (SNR) and uniformity assessments. We aimed to improve imaging setup and coverage and eliminate inter-scan variability by employing multi-slice imaging of a stable phantom. Additionally, we strived to expedite analysis by developing objective, automated analysis software.

methodsA 5300 mL cylindrical plastic bottle placed in plastic bins was scanned at isocenter using a spin-echo sequence with NEMA-recommended parameters and 18 axial slices, avoiding phantom repositioning. Acquisition was repeated with and without prescan normalization filtering and by saving uncombined element images. Obtained data were analyzed using custom open-source MATLAB code. Signal and noise images were automatically assigned, and ROIs for SNR and uniformity calculations were defined using image thresholding. SNR and uniformity pass/fail decisions were made using baseline comparisons.

resultsThe proposed method was successfully implemented as monthly coil QA for 3.5 years. Setup and scanning took 41 min on average for a coil set. Automated image analysis was completed in a few minutes. Signal intensity peaked around +90 or -90 mm for Torso or Head/Neck coil unfiltered images. Noise peaked and minimized SNR inside ±30 mm from isocenter, while maximizing it around ±130 mm. Prescan normalization smoothed signal response, reduced SNR and increased uniformity. Individual coil element image analysis identified their position, signal or noise response and SNR. SNR and uniformity pass/fail thresholds were set for already tested and new coils. Conspicuous and subtle Torso coil malfunctions were detected considering baseline deviations of combined and individual element results.

conclusionsOur QA method eliminated observer bias and provided insights into coil function, image filtering performance and coil element location. It provided SNR and uniformity thresholds and identified faulty coil elements.

Indexed as

HeadMagnetic Resonance ImagingHumansImage Processing, Computer-AssistedPhantoms, ImagingSignal-To-Noise RatioSoftwareMRIdianMR-Linacquality assuranceSNRuniformity

Identifiers

PMID38230873
PMCPMC10930007

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.