Evidence map›Paper›PMID 38223024›Full record

ArticleQuantitative imaging in medicine and surgery2024

Deep learning-based projection synthesis for low-dose cone-beam computed tomography imaging in image-guided radiotherapy.

Xuzhi Zhao, Yi Du, Haizhen Yue, Ruoxi Wang, Shun Zhou, Hao Wu, Wei Wang, Yahui Peng

Abstract read
In one paragraph

Article in Quantitative imaging in medicine and surgery, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed.

  1. Low-dose CBCT image reconstruction: a review.Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al] · 2026
    Review
  2. Article
  3. Article
  4. [Advances in low-dose cone-beam computed tomography image reconstruction methods based on deep learning].Sheng wu yi xue gong cheng xue za zhi = Journal of biomedical engineering = Shengwu yixue gongchengxue zazhi · 2025
    Review
  5. Article
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors.

Xuzhi ZhaoSchool of Electronic and Information Engineering, Beijing Jiaotong University, Beijing, China.
Yi DuKey Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Radiation Oncology, Peking University Cancer Hospital & Institute, Beijing, China.
Haizhen YueKey Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Radiation Oncology, Peking University Cancer Hospital & Institute, Beijing, China.
Ruoxi WangKey Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Radiation Oncology, Peking University Cancer Hospital & Institute, Beijing, China.
Shun ZhouKey Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Radiation Oncology, Peking University Cancer Hospital & Institute, Beijing, China.
Hao WuKey Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Radiation Oncology, Peking University Cancer Hospital & Institute, Beijing, China.
Wei WangSchool of Electronic and Information Engineering, Beijing Jiaotong University, Beijing, China.
Yahui PengSchool of Electronic and Information Engineering, Beijing Jiaotong University, Beijing, China.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: The imaging dose of cone-beam computed tomography (CBCT) in image-guided radiotherapy (IGRT) poses adverse effects on patient health. To improve the quality of sparse-view low-dose CBCT images, a projection synthesis convolutional neural network (SynCNN) model is proposed. Methods: Included in this retrospective, single-center study were 223 patients diagnosed with brain tumours from Beijing Cancer Hospital. The proposed SynCNN model estimated two pairs of orthogonally direction-separable spatial kernels to synthesize the missing projection in between the input neighboring sparse-view projections via local convolution operations. The SynCNN model was trained on 150 real patients to learn patterns for inter-view projection synthesis. CBCT data from 30 real patients were used to validate the SynCNN, while data from a phantom and 43 real patients were used to test the SynCNN externally. Sparse-view projection datasets with 1/2, 1/4, and 1/8 of the original sampling rate were simulated, and the corresponding full-view projection datasets were restored using the SynCNN model. The tomographic images were then reconstructed with the Feldkamp-Davis-Kress algorithm. The root-mean-square error (RMSE), peak signal-to-noise ratio (PSNR), and structural similarity (SSIM) metrics were measured in both the projection and image domains. Five experts were invited to grade the image quality blindly for 40 randomly selected evaluation groups with a four-level rubric, where a score greater than or equal to 2 was considered acceptable image quality. The running time of the SynCNN model was recorded. The SynCNN model was directly compared with the three other methods on 1/4 sparse-view reconstructions. Results: The phantom and patient studies showed that the missing projections were accurately synthesized. In the image domain, for the phantom study, compared with images reconstructed from sparse-view projections, images with SynCNN synthesis exhibited significantly improved qualities with decreased values in RMSE and increased values in PSNR and SSIM. For the patient study, between the results with and without the SynCNN synthesis, the averaged RMSE decreased by 3.4×10 Conclusions: The proposed SynCNN model effectively improves the quality of sparse-view CBCT images at a low time cost. With the SynCNN model, the CBCT imaging dose in IGRT could be reduced potentially.

Indexed as

Cone-beam computed tomography (CBCT)deep learning (DL)low-doseprojection synthesissparse-view

Identifiers

PMID38223024
PMCPMC10784032

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked