Probing Liquid Chromatography-Tandem Mass Spectrometry Response Dynamics and Nonlinear Effects for Response Level Defined Calibration Strategies with Simple Methods To Expand Linear Dynamic Ranges.

Rui Zhang

Open access · goldAbstract read
In one paragraph

Article in ACS omega, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
1.2field-weighted citation impact, top 24% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 8 citations in OpenAlex.

  1. Article
  2. Article
  3. LC-MS/MS Method Validation for Quantification of Nirmatrelvir in Human Plasma.International journal of analytical chemistry · 2025
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

1 author.

Rui ZhangDepartment of Special Chemistry, PathWest Laboratory Medicine, Sir Charles Gardiner Hospital, Nedlands, Western Australia 6009, Australia.ORCID https://orcid.org/0000-0002-2878-3348

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

The response characteristics of liquid chromatography-tandem mass spectrometry (LC-MS/MS) serve as the basis for selecting calibration methods in quantitative analysis. LC-MS/MS inherently exhibits nonlinear detection behavior, primarily attributed to the disproportionate growth observed between peak area and peak height at elevated response levels, potentially leading to signal saturation. This disproportionate peak growth results in reduced unit response (UR), which quantifies the instrument's detection sensitivity. LC-MS/MS typically operates within a narrow near-linear response range (NLRR) due to approximately proportional peak growth, yet the NLRR width varies across different analytes or platforms. Although the inclusion of stable isotope-labeled (SIL) internal standards (IS) in LC-MS/MS analysis can mitigate certain instrument response variations, it does not eliminate the fundamental cause of nonlinearity. Moreover, the concentration range accommodated by the NLRR can significantly fluctuate at different sensitivity levels. LC-MS/MS also encounters various other nonlinear effects, including ion suppression during ionization, signal cross-contribution between the analyte/IS, and matrix effects (ME). Consequently, quadratic regression emerges as a more adaptable approach to LC-MS/MS nonlinear response dynamics, offering a broader calibration range. The application of linear regression, on the other hand, requires strict conditions. Although the signal saturation zone typically remains inaccessible to calibration methods, reducing responses by employing less-optimal selected reaction monitoring (SRM) transitions and/or lower detection gain can facilitate fitting a wide concentration range into the NLRR, thereby enabling accurate linear regression calibration. This report delves into examining the LC-MS/MS response profile, its dynamics, and major nonlinear effects through instrument response mapping to elucidate their influence on the selection of calibration methods.