ArticleJournal of applied clinical medical physics2024
Commissioning and clinical evaluation of a novel high-resolution quality assurance digital detector array for SRS and SBRT.
Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
5 citing papers in PubMed, 7 citations in OpenAlex.
- Clinical assessment of dedicated SRS detector for dose verification in SBRT/SRS treatment plans: to establish clinical threshold with AAPM TG-218 recommendation.Irish journal of medical science · 2026Article
- A multi-institutional EPID-based 3D dose reconstruction model in a virtual phantom for standardized patient QA, benchmarking and auditing for stereotactic radiosurgery.Physics and imaging in radiation oncology · 2026Article
- Patient-specific quality assurance for extremely small field sizes using 3D in vivo dosimetry in fractionated stereotactic radiosurgery for brain metastases.Journal of applied clinical medical physics · 2026Article
- Classification performances of two diode arrays for patient-specific quality assurance of stereotactic body radiation therapy treatments based on absolute dose measurements in phantom.Journal of applied clinical medical physics · 2025Article
- Commissioning and clinical evaluation of a novel high-resolution quality assurance digital detector array for SRS and SBRT.Journal of applied clinical medical physics · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors at 2 institutions in 1 country.
Funding
Abstract
purposeWe aimed to perform the commissioning and clinical evaluation of myQA SRS detector array for patient-specific quality assurance (PSQA) of stereotactic radiosurgery (SRS)/ stereotactic body radiotherapy (SBRT) plans.
methodsTo perform the commissioning of myQA SRS, its dose linearity, dose-rate dependence, angular dependence, and field-size dependence were investigated. Ten SBRT plans were selected for clinical evaluation: 1) Common clinical deviations based on the original SBRT plan (Plan0), including multileaf collimator (MLC) positioning deviation and treatment positioning deviation were introduced. 2) Compared the performance of the myQA SRS and a high-resolution EPID dosimetry system in PSQA measurement for the SBRT plans. Evaluation parameters include gamma passing rate (GPR) and distance-to-agreement (DTA) pass rate (DPR).
resultsThe dose linearity, angle dependence, and field-size dependence of myQA SRS system exhibit excellent performance. The myQA SRS is highly sensitive in the detection of MLC deviations. The GPR of (3%/1 mm) decreases from 90.4% of the original plan to 72.7%/62.9% with an MLC outward/inward deviation of 3 mm. Additionally, when the setup error deviates by 1 mm in the X, Y, and Z directions with the GPR of (3%/1 mm) decreasing by an average of -20.9%, -25.7%, and -24.7%, respectively, and DPR (1 mm) decreasing by an average of -33.7%, -32.9%, and -29.8%. Additionally, the myQA SRS has a slightly higher GPR than EPID for PSQA, However, the difference is not statistically significant with the GPR of (3%/1 mm) of (average 90.4%% vs. 90.1%, p = 0.414).
conclusionDosimetry characteristics of the myQA SRS device meets the accuracy and sensitivity requirement of PSQA for SRS/SBRT treatment. The dose rate dependence should be adequately calibrated before its application and a more stringent GPR (3%/1 mm) evaluation criterion is suggested when it is used for SRS/SBRT QA.