A practical method for assessing quantitative scanner accuracy with long-lived radionuclides: The ARTnet insert.

Dale L Bailey, Kathy P Willowson, Carl Muñoz-Ferrada

Open access · greenAbstract read
In one paragraph

Article in Asia Oceania journal of nuclear medicine & biology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
2.0field-weighted citation impact, top 15% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 5 citations in OpenAlex.

  1. Standardization of PET/CT Performance Requirements for Whole-Body Quantitative Imaging: An International Proposal.Journal of nuclear medicine : official publication, Society of Nuclear Medicine · 2025
    Review
  2. ARTnet Perspectives and Contributions to Theranostics.World journal of nuclear medicine · 2025
    Review
  3. Dosimetry Software for Theranostic Applications: Current Capabilities and Future Prospects.Journal of nuclear medicine : official publication, Society of Nuclear Medicine · 2025
    Review
  4. 360° CZT-SPECT/CT cameras:EJNMMI physics · 2024
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors at 2 institutions in 1 country.

Dale L BaileyDepartment of Nuclear Medicine, Royal North Shore Hospital, Sydney, Australia.
Kathy P WillowsonDepartment of Nuclear Medicine, Royal North Shore Hospital, Sydney, Australia.
Carl Muñoz-FerradaGlobalsonics Institute for Medical Research, Sydney, Australia.
Royal North Shore Hospital · AUUniversity of Sydney · AU

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Objectives: To address the problem of using large volumes of long-lived radionuclides in test phantoms to check calibration accuracy of PET and SPECT systems we have developed a test object which (a) contains less radioactivity, (b) has a low total volume, and (c) is easier to store than currently used phantoms, while still making use of readily-available "standardised" test objects. Methods: We have designed a hollow acrylic cylindrical insert compatible with the NEMA/IEC PET Body Image Quality (IQ) phantom used in NU 2 performance testing of PET systems. The insert measures 90 mm internal diameter and 70 mm internal height and so is sufficiently large to not be subject to partial volume effects in PET or SPECT imaging. The volume of the insert is approximately 500 mL. It has been designed as a replacement for the standard long cylindrical "lung insert" in the IQ phantom without needing to remove the fillable hollow spheres of the phantom. The insert been tested with Results: The scanners were found to produce accurate image reconstructions in the insert with  5% of the true value without any confounding uncertainty from partial volume effects when compared to NEMA NU 2-2018 Phantom measurement. Conclusions: The "ARTnet Insert" is simple to use, inexpensive, compatible with current phantoms and is suitable for both PET and SPECT systems. It does not suffer from significant partial volume losses permitting its use even with the poor spatial resolution of high-energy imaging with

Indexed as

Clinical TrialPETSite validationSPECT

Identifiers

PMID38164228
PMCPMC10757053
OpenAlexW4390507611