ArticleJournal of applied clinical medical physics2024
Dosimetric effects of rotational errors for single isocenter multiple targets in HyperArc plans: A phantom and retrospective imaging analysis study.
Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
5 citing papers in PubMed.
- Assessing the role of plan complexity and target geometry through multi-institutional gel based end-to-end QA in multi-focal single isocenter stereotactic radiosurgery.Journal of applied clinical medical physics · 2026Article
- Article
- Evaluation of artificial intelligence and optical image recognition techniques used in OneIso, an off-axis Winston-Lutz quality assurance phantom.Biomedical physics & engineering express · 2024Article
- Dosimetric effects of rotational errors for single isocenter multiple targets in HyperArc plans: A phantom and retrospective imaging analysis study.Journal of applied clinical medical physics · 2024Article
- A multi-institutional trial evaluating the use of an integrated quality assurance phantom for frameless single-isocenter multitarget stereotactic radiosurgery.Frontiers in oncology · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThis study uses a phantom to investigate the dosimetric impact of rotational setup errors for Single Isocenter Multiple Targets (SIMT) HyperArc plans. Additionally, it evaluates intra-fractional rotational setup errors in patients treated with Encompass immobilization system.
methodsThe Varian HyperArc system (Varian Medical systems) was used to create plans targeting spherical PTVs with diameters of 5, 10, and 15 mm and with offsets of 1.3-5.3 cm from the isocenter. Dosimetric parameters, including mean and maximum dose, D99% and D95% were evaluated for various rotational setup errors ranging from 0.5° to 2° for the PTVs and certain CTVs created within PTVs. These rotational errors were applied in an order and direction that resulted in the maximum displacement of targets. The rotation was applied both uniformly around all three axes and individually around each axis. Furthermore, to link the findings to actual treatment scenarios, the intra-fractional rotational setup errors were obtained for stereotactic cranial patients treated with the Encompass system using CBCT images acquired during treatments.
resultsThe maximum displacement of 2.7 mm was observed for targets located at 4.4 and 4.5 cm from the isocenter with rotational setup errors of 2°. The dose reduction for D99% values corresponding to this displacement were about 50%, 40%, and 30% for PTVs with diameters of 5, 10, and 15 mm, respectively. Both D99% and D95% showed a consistent trend of dose reduction across various rotational errors and PTV volumes. While the maximum dose remained consistent for different targets with various rotational errors, the mean dose decreased by approximately 25%, 12%, and 6% for PTVs with diameters of 5, 10, and 15 cm, respectively, with rotational errors of 2°. In addition, by analyzing CBCT images, the absolute mean rotational setup errors obtained during treatment with Encompass for pitch, roll, and yaw were 0.17° ± 0.13°, 0.11° ± 0.10°, and 0.12° ± 0.10° respectively. This data, combined with existing studies, suggest that a 0.5° rotational setup error is a safe choice to consider for calculating additional PTV margin to ensure adequate CTV coverage. Therefore, the assessment of maximum displacement and dosimetric parameters in this study, for a 0.5° rotational error, highlights the need for an additional 0.7 mm PTV margin for targets positioned at distances of 4.4 cm or greater from the isocenter.
conclusionsFor SIMT Plans, a 0.5° rotational setup error is recommended as a basis for calculating additional PTV margins to ensure adequate CTV coverage when using the Encompass system.