ArticleDiabetes therapy : research, treatment and education of diabetes and related disorders2024

Evaluation of System Accuracy, Precision, Hematocrit Influence, and User Performance of Two Blood Glucose Monitoring Systems Based on ISO 15197:2013/EN ISO 15197:2015.

Stefan Pleus, Nina Jendrike, Annette Baumstark, Jochen Mende, Stephanie Wehrstedt, Cornelia Haug, Guido Freckmann

Abstract read
In one paragraph

Article in Diabetes therapy : research, treatment and education of diabetes and related disorders, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Stefan PleusInstitut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Lise-Meitner-Strasse 8/2, 89081, Ulm, Germany.
Nina JendrikeInstitut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Lise-Meitner-Strasse 8/2, 89081, Ulm, Germany.
Annette BaumstarkInstitut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Lise-Meitner-Strasse 8/2, 89081, Ulm, Germany.
Jochen MendeInstitut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Lise-Meitner-Strasse 8/2, 89081, Ulm, Germany.
Stephanie WehrstedtInstitut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Lise-Meitner-Strasse 8/2, 89081, Ulm, Germany. stephanie.wehrstedt@idt-ulm.de.ORCID http://orcid.org/0009-0000-4586-4660
Cornelia HaugInstitut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Lise-Meitner-Strasse 8/2, 89081, Ulm, Germany.
Guido FreckmannInstitut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Lise-Meitner-Strasse 8/2, 89081, Ulm, Germany.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionSufficiently high analytical quality of blood glucose monitoring systems (BGMS) is a prerequisite for efficient diabetes therapy. In this study we assessed system accuracy, measurement repeatability, intermediate measurement precision, user performance, and the influence of hematocrit on two CE-marked blood glucose monitoring systems. For one BGMS, measurement accuracy using venous samples was additionally investigated.

methodsStudy procedures were based on the International Organization of Standardization (ISO) 15197:2013/EN ISO 15197:2015 ("ISO 15197"). User performance included data from 100 subjects who used one test strip lot, whereas for all other analyses three different reagent system lots were used. For system accuracy assessment, 100 capillary samples were measured in duplicate with each of three reagent system lots per system, resulting in 600 results per system.

resultsCareSens S Fit and CareSens H Beat both fulfilled the ISO 15197 accuracy criteria with 97.5-100% of each test strip lot's results falling within ± 15 mg/dL or ± 15% of the results of the comparison method and 100% of results in consensus error grid (CEG) zone A for all three lots. User performance evaluation revealed sufficient accuracy in the hands of lay users although some handling errors were documented by study staff. Assessment of measurement repeatability and intermediate measurement precision is given by standard deviation (SD) (glucose levels < 100 mg/dL) and by coefficient of variation (CV) (glucose concentrations ≥ 100 mg/dL). SD was ≤ 4.1 mg/dL and CV ≤ 4.2% for measurement repeatability and SD was ≤ 2.2 mg/dL and CV ≤ 2.6% for intermediate measurement precision. In case of hematocrit influence, both BGMS complied with all three tested lots with the defined criteria.

conclusionBoth BGMS analyzed in this study fulfilled the required accuracy criteria of ISO 15197. They showed high precision, good performance in the hands of lay users, and the influence of hematocrit was acceptable in the labeled range.

Indexed as

Blood glucose monitoring systemEN ISO 15197:2015Hematocrit influenceIntermediate measurement repeatabilityISO 15197:2013Measurement precisionSystem accuracyUser performance

Identifiers

PMID38100029
PMCPMC10838869

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.