Evidence map›Paper›PMID 38075549›Full record

ArticleMedical devices (Auckland, N.Z.)2023

A Comparison of the Preclinical Performance of the Echelon™+ Stapler with Thunderbird Reloads to Two Commercial Endoscopic Surgical Staplers.

Shanshan Wang, Ying Hua, Jun Liu, Zhifan F Huang, Jeffrey W Clymer, Crystal D Ricketts, Jin Hao

Abstract read
In one paragraph

Article in Medical devices (Auckland, N.Z.), 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Shanshan WangEthicon, Inc, Cincinnati, OH, USA.
Ying HuaEthicon, Inc, Cincinnati, OH, USA.
Jun LiuEthicon, Inc, Cincinnati, OH, USA.
Zhifan F HuangEthicon, Inc, Cincinnati, OH, USA.
Jeffrey W ClymerEthicon, Inc, Cincinnati, OH, USA.ORCID 0000-0001-9472-6192
Crystal D RickettsEthicon, Inc, Cincinnati, OH, USA.
Jin HaoEthicon, Inc, Cincinnati, OH, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: Design of surgical staplers continues to advance with more consistent staple formation that can lead to higher leak pressures and lower rates of leak along the staple line. This study was performed to compare the Ethicon Echelon™+ Stapler with Thunderbird reloads to two other currently available commercial staplers, Reach Anzhi and Fulbright Lunar with corresponding reloads. Methods: The rate of malformed staples for three staplers was determined in porcine stomach (3.0-3.3 mm thickness) via CT scanning. Staple line air leak pressures in bronchial tissue (3.0-3.3 mm) and fluid leak pressures in colon (1.3-1.7 mm) were measured and compared to a standard success criterion for both tissues. Results: The rate of malformed staples in gastric tissue for Echelon+ was more than 90% lower than for the two other commercial staplers (p < 0.001). In bronchus, Echelon+ had 56% higher air leak pressures than Reach Anzhi (p < 0.001) and was not significantly different from Fulbright Lunar. In colon, Echelon+ had over twice the fluid leak pressures of the comparators (p < 0.001). Conclusion: The Echelon+ Stapler with Thunderbird reloads exhibited a low rate of malformed staples, and its staple lines withstood high leak pressures in both thick and thin tissues. Clinical studies are needed to confirm that these observed benefits carry over into actual practice.

Indexed as

burst pressureendoscopicleakingmalformed staplesstapler

Identifiers

PMID38075549
PMCPMC10710239

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.