Evidence map›Paper›PMID 38028485›Full record

ArticleFrontiers in cardiovascular medicine2023

Remotely detuned receiver coil for high-resolution interventional cardiac magnetic resonance imaging.

Sina Marhabaie, Marylène Delcey, Dounia El Hamrani, Fanny Vaillant, Jean-Christophe Ginefri, Valéry Ozenne, Emma Abell, Marie Poirier-Quinot, Bruno Quesson

Open access · goldAbstract read
In one paragraph

Article in Frontiers in cardiovascular medicine, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
0.2field-weighted citation impact, top 39% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed, 1 citations in OpenAlex.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors at 6 institutions in 1 country.

Sina Marhabaie *Laboratoire D'Imagerie Biomédicale Multimodale Paris Saclay, Université Paris-Saclay, CNRS, Inserm, Orsay, France.
Marylène Delcey *Univ. Bordeaux, INSERM, CRCTB, U 1045, IHU Liryc, Bordeaux, France.
Dounia El HamraniUniv. Bordeaux, INSERM, CRCTB, U 1045, IHU Liryc, Bordeaux, France.
Fanny VaillantUniv. Bordeaux, INSERM, CRCTB, U 1045, IHU Liryc, Bordeaux, France.
Jean-Christophe GinefriLaboratoire D'Imagerie Biomédicale Multimodale Paris Saclay, Université Paris-Saclay, CNRS, Inserm, Orsay, France.
Valéry OzenneUniv. Bordeaux, CNRS, CRMSB, UMR 5536, IHU Liryc, Bordeaux, France.
Emma AbellUniv. Bordeaux, INSERM, CRCTB, U 1045, IHU Liryc, Bordeaux, France.
Marie Poirier-QuinotLaboratoire D'Imagerie Biomédicale Multimodale Paris Saclay, Université Paris-Saclay, CNRS, Inserm, Orsay, France.
Bruno Quesson *Univ. Bordeaux, CNRS, CRMSB, UMR 5536, IHU Liryc, Bordeaux, France.
Université de Bordeaux · FRUniversité Paris-Saclay · FRCentre de Recherche Cardio-Thoracique de Bordeaux · FRCentre National de la Recherche Scientifique · FRElectrophysiology and Heart Modeling Institute · FRSiemens (France) · FR

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Introduction: Interventional cardiac MRI in the context of the treatment of cardiac arrhythmia requires submillimeter image resolution to precisely characterize the cardiac substrate and guide the catheter-based ablation procedure in real-time. Conventional MRI receiver coils positioned on the thorax provide insufficient signal-to-noise ratio (SNR) and spatial selectivity to satisfy these constraints. Methods: A small circular MRI receiver coil was developed and evaluated under different experimental conditions, including high-resolution MRI anatomical and thermometric imaging at 1.5 T. From the perspective of developing a therapeutic MR-compatible catheter equipped with a receiver coil, we also propose alternative remote active detuning techniques of the receiver coil using one or two cables. Theoretical details are presented, as well as simulations and experimental validation. Results: Anatomical images of the left ventricle at 170 µm in-plane resolution are provided on Discussion: This study shows the potential of small dimension surface coils for minimally invasive therapy of cardiac arrhythmia intraoperatively guided by MRI. The proposed remote decoupling approaches may simplify the construction process and reduce the cost of such single-use devices.

Indexed as

cathetercoil designdetuningheartinterventionMRIMR-thermometry

Identifiers

PMID38028485
PMCPMC10643167
OpenAlexW4388096651

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.