ArticleScientific reports2023

Comparing the biomechanical properties of conventional suture and all-suture anchors using patient-specific and realistic osteoporotic and non-osteoporotic phantom using 3D printing.

Junhyeok Ock, Junghyun Seo, Kyoung Hwan Koh, Namkug Kim

Open access · goldAbstract read
In one paragraph

Article in Scientific reports, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed
3.5field-weighted citation impact, top 7% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

8 citing papers in PubMed, 11 citations in OpenAlex.

  1. Review
  2. Review
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. 3D Bioprinting in Limb Salvage Surgery.Journal of functional biomaterials · 2024
    Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors at 2 institutions in 1 country.

Junhyeok OckDepartment of Biomedical Engineering, Asan Medical Center, Asan Medical Institute of Convergence Science and Technology, University of Ulsan College of Medicine, Pungnap2-dong, Songpa-gu, Seoul, South Korea.
Junghyun SeoDepartment of Orthopaedic Surgery, Asan Medical Center, University of Ulsan College of Medicine, 88 Olympic-Ro 43-Gil Songpa-Gu, Seoul, 05505, South Korea.
Kyoung Hwan Koh *Department of Orthopaedic Surgery, Asan Medical Center, University of Ulsan College of Medicine, 88 Olympic-Ro 43-Gil Songpa-Gu, Seoul, 05505, South Korea. osdoc.koh@gmail.com.
Namkug Kim *Department of Biomedical Engineering, Asan Medical Center, Asan Medical Institute of Convergence Science and Technology, University of Ulsan College of Medicine, Pungnap2-dong, Songpa-gu, Seoul, South Korea. namkugkim@gmail.com.
Asan Medical Center · KRUlsan College · KR

Funding

Korea Health Industry Development Institute HR20C0026
6 · The paper itself

Abstract

Conventional suture anchors (CAs) and all-suture anchors (ASAs) are used for rotator cuff repair. Pull-out strength (POS) is an important factor that affects surgical outcomes. While the fixation mechanism differs between the anchor types and relies on the quality, few studies have compared biomechanical properties of anchors based on bone quality. This study aimed to compare the biomechanical properties of anchors using osteoporotic bone (OB) and non-osteoporotic bone (NOB) simulators. Humerus simulators were fabricated using fused deposition modeling of 3D printing and acrylonitrile butadiene styrene adjusting the thickness of cortical bone and density of cancellous bone based on CT images. Cyclic loading from 10 to 50 N, 10 to 100 N, and 10 to 150 N for 10 cycles was clinically determined at each anchor because the supraspinatus generates a force of 67-125 N in daily activities of normal control. After cyclic loading, the anchor was extruded at a load of 5 mm/min. Displacement, POS, and stiffness were measured. In OB simulators, CAs revealed bigger gap displacement than ASAs with cyclic loading of 10-150 N. ASA showed higher values for POS and stiffness. In NOB simulators, ASAs revealed bigger gap displacement than CAs with cyclic loading of 10-150 N. ASA showed higher values for POS and CA showed higher values for stiffness. POS of anchors depends on anchors 'displacement and bone stiffness. In conclusion, ASA demonstrated better biomechanical performance than CA in terms of stability under cyclic loading and stiffness with similar POS in OB.

Indexed as

Rotator Cuff InjuriesSuture AnchorsBiomechanical PhenomenaCadaverHumansHumerusPrinting, Three-DimensionalSuturesSuture Techniques

Identifiers

PMID38017064
PMCPMC10684536
OpenAlexW4389084909

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.