The performance of a new type accelerator uRT-linac 506c evaluated by a quality assurance automation system.

WenZhao Sun, ZhongHua Shi, Xin Yang, SiJuan Huang, Can Liao, Wei Zhang, YongBao Li, XiaoYan Huang

Open access · goldAbstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
2.1field-weighted citation impact, top 13% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 6 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 4 institutions in 1 country.

WenZhao SunState Key Laboratory of Oncology in South China, Collaborative Innovation Center for Cancer Medicine, Sun Yat-sen University Cancer Center, Guangzhou, China.
ZhongHua ShiRadiotherapy and Imaging R&D department, Shanghai United Imaging Healthcare Co., Ltd., Shanghai, China.
Xin YangState Key Laboratory of Oncology in South China, Collaborative Innovation Center for Cancer Medicine, Sun Yat-sen University Cancer Center, Guangzhou, China.
SiJuan HuangState Key Laboratory of Oncology in South China, Collaborative Innovation Center for Cancer Medicine, Sun Yat-sen University Cancer Center, Guangzhou, China.
Can LiaoRadiotherapy and Imaging R&D department, Shanghai United Imaging Healthcare Co., Ltd., Shanghai, China.
Wei ZhangRadiotherapy and Imaging R&D department, Shanghai United Imaging Healthcare Co., Ltd., Shanghai, China.
YongBao LiState Key Laboratory of Oncology in South China, Collaborative Innovation Center for Cancer Medicine, Sun Yat-sen University Cancer Center, Guangzhou, China.
XiaoYan HuangState Key Laboratory of Oncology in South China, Collaborative Innovation Center for Cancer Medicine, Sun Yat-sen University Cancer Center, Guangzhou, China.
United Imaging Healthcare (China) · CNSun Yat-sen University · CNSun Yat-sen University Cancer Center · CNGuangzhou Medical University · CN

Funding

National Natural Science Foundation of China 12275372National Natural Science Foundation of China 82373528Science and technology project of Guangdong Esophageal Cancer Institute Q202008
6 · The paper itself

Abstract

purposeThe purpose of this study was to evaluate the performance of our quality assurance (QA) automation system and to evaluate the machine performance of a new type linear accelerator uRT-linac 506c within 6 months using this system.

methodsThis QA automation system consists of a hollow cylindrical phantom with 18 steel balls in the phantom surface and an analysis software to process electronic portal imaging device (EPID) measurement image data and report the results. The performance of the QA automation system was evaluated by the tests of repeatability, archivable precision, detectability of introduced errors, and the impact of set-up errors on QA results. The performance of this linac was evaluated by 31 items using this QA system over 6 months.

resultsThis QA system was able to automatically deliver QA plan, EPID image acquisition, and automatic analysis. All images acquiring and analysis took approximately 4.6 min per energy. The preset error of 0.1 mm in multi-leaf collimator (MLC) leaf were detected as 0.12 ± 0.01 mm for Bank A and 0.10 ± 0.01 mm in Bank B. The 2 mm setup error was detected as -1.95 ± 0.01 mm, -2.02 ± 0.01 mm, 2.01 ± 0.01 mm for X, Y, Z directions, respectively. And data from the tests of repeatability and detectability of introduced errors showed the standard deviation were all within 0.1 mm and 0.1°. and data of the machine performance were all within the tolerance specified by AAPM TG-142.

conclusionsThe QA automation system has high precision and good performance, and it can improve the QA efficiency. The performance of the new accelerator has also performed very well during the testing period.

Indexed as

Particle AcceleratorsRadiotherapy, Intensity-ModulatedAutomationHumansPhantoms, ImagingQuality Assurance, Health CareRadiotherapy Planning, Computer-AssistedSoftwareEPIDmachine performancemulti-leaf collimatorquality assurance automation