Optimization of Y-90 Radioembolization Imaging for Post-Treatment Dosimetry on a Long Axial Field-of-View PET/CT Scanner.
Article in Diagnostics (Basel, Switzerland), 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
11 citing papers in PubMed, 12 citations in OpenAlex.
- Large axial field of view PET/CT - Not just a longer PET gantry.Zeitschrift fur medizinische Physik · 2026Review
- Impact of scan duration and scatter correction on quantitative accuracy in Yttrium-90 PET/CT imaging using LAFOV and SAFOV systems: a phantom and clinical evaluation.EJNMMI physics · 2026Article
- Continuous bed motion versus step-and-shoot acquisition in LAFOV PET/CT: insights from multi-phantom and patient studies.EJNMMI physics · 2026Article
- Multimodal imaging techniques in Yttrium-90 radioembolization for hepatocellular carcinoma: a modality-oriented clinical review.Frontiers in oncology · 2026Review
- Long Axial Field-of-view PET: A New Era of Quantitative PET Imaging for Theranostic Applications in Clinic.PET clinics · 2025Review
- Theranostic Radioembolization: Radiation Dosimetry-Guided Treatment Planning and Delivery.World journal of nuclear medicine · 2025Review
- A digital twin of the Biograph Vision Quadra long axial field of view PET/CT: Monte Carlo simulation and image reconstruction framework.EJNMMI physics · 2025Article
- Current Research in Drug-Free Cancer Therapies.Bioengineering (Basel, Switzerland) · 2025Review
- Low-activity [EJNMMI physics · 2025Article
- Y-90 PET/MR imaging optimization with a Bayesian penalized likelihood reconstruction algorithm.Physical and engineering sciences in medicine · 2024Article
- Optimization and impact of sensitivity mode on abbreviated scan protocols with population-based input function for parametric imaging of [European journal of nuclear medicine and molecular imaging · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors at 2 institutions in 1 country.
Funding
Abstract
backgroundPET imaging after yttrium-90 (Y-90) radioembolization is challenging because of the low positron fraction of Y-90 (32 × 10
methodsTwo phantoms (NEMA IEC and AbdoMan phantoms, mimicking human liver) filled with Y-90 and a 4:1 sphere (tumor)-to-background ratio were scanned for 24 h with the Biograph Vision Quadra (Siemens Healthineers). Eight patients were scanned after Y-90 radioembolization (1.3-4.7 GBq) using the optimized protocol (obtained by phantom studies). The IQ, contrast recovery coefficients (CRCs) and noise were evaluated for their limited and full acceptance angles, different rebinned scan durations, numbers of iterations and post-reconstruction filters. The s-value-based absorbed doses were calculated to assess their suitability for dosimetry.
resultsThe phantom studies demonstrate that two iterations, five subsets and a 4 mm Gaussian filter provide a reasonable compromise between a high CRC and low noise. For a 20 min scan duration, an adequate CRC of 56% (vs. 24 h: 62%, 20 mm sphere) was obtained, and the noise was reduced by a factor of 1.4, from 40% to 29%, using the full acceptance angle. The patient scan results were consistent with those from the phantom studies, and the impacts on the absorbed doses were negligible for all of the studied parameter sets, as the maximum percentage difference was -3.89%.
conclusionsWith 2i5s, a 4 mm filter and a scan duration of 20 min, IQ and quantification accuracy that are suitable for post-treatment dosimetry of Y-90 radioembolization can be achieved.