ArticleJournal of applied clinical medical physics2024
Characteristics of a plastic scintillation detector in photon beam dosimetry.
Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed.
- Performance evaluation of a new type of commercial silicon detectors for photon beam dosimetry.Journal of applied clinical medical physics · 2026Article
- Pulse‑level characterization of low monitor unit deliveries on a modern linear accelerator using a plastic scintillation detector.Journal of applied clinical medical physics · 2026Article
- Characterization of a novel plastic scintillation detector for in vivo electron dosimetry.Journal of applied clinical medical physics · 2026Article
- Ultra-fast dosimetric data collection with a commercial plastic scintillation detector in an MR-linac.Journal of applied clinical medical physics · 2026Article
- Fast and accurate measurement of small field dosimetry using a novel scintillation detector.Journal of applied clinical medical physics · 2025Article
- A novel Cherenkov radiation removal method for plastic scintillator detectors in a 0.35 T MR-Linac.Journal of applied clinical medical physics · 2025Article
- Grid/lattice therapy: consideration of small field dosimetry.The British journal of radiology · 2024Review
- Characteristics of a plastic scintillation detector in photon beam dosimetry.Journal of applied clinical medical physics · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundPlastic scintillating detectors (PSD) have gained popularity due to small size and are ideally suited in small-field dosimetry due to no correction needed and hence detector reading can be compared to dose. Likewise, these detectors are active and water equivalent. A new PSD from Blue Physics is characterized in photon beam. PURPOSE: Innovation in small-field dosimetry detector has led us to examine Blue Physics PSD (BP-PSD) for use in photon beams from linear accelerator.
methodsBP-PSD was acquired and its characteristics were evaluated in photon beams from a Varian TrueBeam. Data were collected in a 3D water tank. Standard parameters; dose, dose rate, energy, angular dependence and temperature dependence were studied. Depth dose, profiles and output in a reference condition as well as small fields were measured.
resultsBP-PSD is versatile and provides data very similar to an ion chamber when Cerenkov radiation is properly accounted. This device measures data pulse by pulse which very few detectors can perform. The differences between ion chamber data and PSD are < 2% in most cases. The angular dependence is a bit pronounces to 1.5% which is due to PSD housing. Depth dose and profiles are comparable within < 1% to an ion chamber. For small fields this detector provides suitable field output factor compared to other detectors and Monte Carlo (MC) simulated data without any added correction factor.
conclusionsThe characteristics of Blue Physics PSD is uniquely suitable in photon beam and more so in small fields. The data are reproducible compared to ion chamber for most parameters and ideally suitable for small-field dosimetry without any correction factor.