A simplified low-cost phantom for image quality assessment of dental cone beam computed tomography unit.
Article in Journal of medical radiation sciences, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Design and Use of a Custom Phantom for Regular Tests of Radiography Apparatus: A Feasibility Study.Journal of imaging · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionA standardised testing protocol for evaluation of a wide range of dental cone beam computed tomography (CBCT) performance and image quality (IQ) parameters is still limited and commercially available testing tool is unaffordable by some centres. This study aims to assess the performance of a low-cost fabricated phantom for image quality assessment (IQA) of digital CBCT unit.
methodsA customised polymethyl methacrylate (PMMA) cylindrical phantom was developed for performance evaluation of Planmeca ProMax 3D Mid digital dental CBCT unit. The fabricated phantom consists of four different layers for testing specific IQ parameters such as CT number accuracy and uniformity, noise and CT number linearity. The phantom was scanned using common scanning protocols in clinical routine (90.0 kV, 8.0 mA and 13.6 s). In region-of-interest (ROI) analysis, the mean CT numbers (in Hounsfield unit, HU) and noise for water and air were determined and compared with the reference values (0 HU for water and -1000 HU for air). For linearity test, the correlation between the measured HU of different inserts with their density was studied.
resultsThe average CT number were -994.1 HU and -2.4 HU, for air and water, respectively and the differences were within the recommended acceptable limit. The linearity test showed a strong positive correlation (R
conclusionThe fabricated IQ phantom serves as a simple and affordable testing tool for digital dental CBCT imaging.