Evidence map›Paper›PMID 37962024›Full record

ArticleJournal of applied clinical medical physics2023

Technical note: Commissioning of a low-cost system for directly 3D printed flexible bolus.

Garrett C Baltz, Steven M Kirsner

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors.

Garrett C BaltzScripps Cancer Center, San Diego, California, USA.ORCID https://orcid.org/0000-0002-3967-1026
Steven M KirsnerScripps Cancer Center, San Diego, California, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeTo present the commissioning process of a low-cost solution for directly 3D printed flexible patient specific bolus.

methodsThe 3D printing solution used in this study consisted of a resin stereolithography 3D printer and a flexible curing resin. To test the dimensional accuracy of the 3D printer, rectangular cuboids with varying dimensions were 3D printed and their measured dimensions were compared to the designed dimensions. Percent Depth Dose (PDD) profiles were measured by irradiating film embedded in a 3D printed phantom made of the flexible material. A CT of the phantom was acquired and used to replicate the irradiation setup in the treatment planning system. PDDs were calculated for both the native HU of the phantom, and with the phantom HU overridden to 300 HU to match its physical density. Dosimetric agreement was characterized by comparing calculated to measured depths of R90, R80, and R50. Upon completion of the commissioning process, a bolus was 3D printed for a clinical case study for treatment of the nose.

resultsDimensional accuracy of the printer and material combination was found to be good, with all measured dimensions of test cuboids within 0.5 mm of designed. PDD measurements demonstrated the best dosimetric agreement when the material was overridden to 300 HU, corresponding to the measured physical density of the material of 1.18 g/cc. Calculated and measured depths of R90, R80, and R50 all agreed within 1 mm. The bolus printed for the clinical case was free from defects, highly conformal, and led to a clinically acceptable plan.

conclusionThe results of the commissioning measurements performed indicate that the 3D printer and material solution are suitable for clinical use. The 3D printer and material combination can provide a low-cost solution a clinic can implement in-house to directly 3D print flexible bolus.

Indexed as

NoseRadiometryHumansPhantoms, ImagingPrinting, Three-DimensionalRadiotherapy Dosage3D printingbolusdosimetryelectron radiotherapytreatment planning

Identifiers

PMID37962024
PMCPMC10691640

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.