Comparison of baseline drifts using three reflector blocks versus using a single reflector block for the calibration of wall-mounted Respiratory Gating for Scanner (RGSC) camera integrated with a CT.
Article in Journal of applied clinical medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
6 citing papers in PubMed.
- A novel respiratory surrogate system with table motion correction and visual feedback for computed tomography.Physics and imaging in radiation oncology · 2025Article
- A self-gated 4D-MRI sequence for internal target volume delineation in liver: Phantom and pre-clinical validation.Journal of applied clinical medical physics · 2025Article
- Investigating the effects of table movement and sag on optical surrogate-driven respiratory-guided computed tomography.Journal of applied clinical medical physics · 2025Article
- Enhanced calibration method for wall-mounted RGSC cameras: Mitigating baseline drift caused by CT couch sagging.PloS one · 2025Article
- Resolving signal drift in the wall-mounted camera of the RGSC system.Journal of applied clinical medical physics · 2024Article
- Comparison of baseline drifts using three reflector blocks versus using a single reflector block for the calibration of wall-mounted Respiratory Gating for Scanner (RGSC) camera integrated with a CT.Journal of applied clinical medical physics · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThe calibration of the Respiratory Gating for SCanner (RGSC) system is critical to achieve better and more stable accuracy. The current procedure for a wall-mounted RGSC system has a relatively large residual error. PURPOSE: To compare the baseline drifts in the image acquisition of DIBH using three reflector blocks versus using a single reflector block in the calibration of a wall-mounted RGSC camera system. MATERIALS AND
methodsVarian provides a calibration plate with three rows of calibration points: each row is separated by 15 cm longitudinally and by 10 cm laterally. In Varian's single-block calibration method, the reflector block was first placed on the center point of the calibration plate and aligned with the scanner isocenter. The calibration took a picture of the block, then placed the block on the other eight points sequentially. In the proposed three-block method, we placed three reflector blocks on the center row, with the center block aligned with the isocenter, and we took a picture of the center block by manually blocking the other two blocks in calibration. By moving the couch longitudinally in or out 15 cm, the calibration goes through all nine points. Monte Carlo simulation was done using Matlab to analyze the calibration matrix eigenvalue characteristics.
resultsFor a typical scan length of 40 cm of DIBH, the residual baseline drift in simulated DIBH is 0.02 ± 0.03 versus 0.30 ± 0.12 cm for three-block calibration and single-block calibration, respectively. To achieve 0.5 mm tolerance for the eigenvalue, the laser and reflector box should be within ±3 mm uncertainties based on the eigenvalue simulation.
conclusionThree-block calibration method effectively removes baseline drift caused by couch movement in DIBH/4D CT scan for the wall-mounted camera while the single-block calibration method still has significant residual baseline drift.