CD30 Lateral Flow and Enzyme-Linked Immunosorbent Assays for Detection of BIA-ALCL: A Pilot Study.
Article in Cancers, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed, 4 citations in OpenAlex.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
13 authors at 6 institutions in 2 countries.
Funding
Abstract
introductionBreast Implant-Associated Anaplastic Large Cell Lymphoma (BIA-ALCL) commonly presents as a peri-implant effusion (seroma). CD30 (TNFRSF8) is a consistent marker of tumor cells but also can be expressed by activated lymphocytes in benign seromas. Diagnosis of BIA-ALCL currently includes cytology and detection of CD30 by immunohistochemistry or flow cytometry, but these studies require specialized equipment and pathologists' interpretation. We hypothesized that a CD30 lateral flow assay (LFA) could provide a less costly rapid test for soluble CD30 that eventually could be used by non-specialized personnel for point-of-care diagnosis of BIA-ALCL.
methodsWe performed LFA for CD30 and enzyme-linked immunosorbent assay (ELISA) for 15 patients with pathologically confirmed BIA-ALCL and 10 patients with benign seromas. To determine the dynamic range of CD30 detection by LFA, we added recombinant CD30 protein to universal buffer at seven different concentrations ranging from 125 pg/mL to 10,000 pg/mL. We then performed LFA for CD30 on cryopreserved seromas of 10 patients with pathologically confirmed BIA-ALCL and 10 patients with benign seromas.
resultsRecombinant CD30 protein added to universal buffer produced a distinct test line at concentrations higher than 1000 pg/mL and faint test lines at 250-500 pg/mL. LFA produced a positive test line for all BIA-ALCL seromas undiluted and for 8 of 10 malignant seromas at 1:10 dilution, whereas 3 of 10 benign seromas were positive undiluted but all were negative at 1:10 dilution. Undiluted CD30 LFA had a sensitivity of 100.00%, specificity of 70.00%, positive predictive value of 76.92%, and negative predictive value of 100.00% for BIA-ALCL. When specimens were diluted 1:10, sensitivity was reduced to 80.00% but specificity and positive predictive values increased to 100.00%, while negative predictive value was reduced to 88.33%. When measured by ELISA, CD30 was below 1200 pg/mL in each of six benign seromas, whereas seven BIA-ALCL seromas contained CD30 levels > 2300 pg/mL, in all but one case calculated from dilutions of 1:10 or 1:50.
conclusionsBIA-ALCL seromas can be distinguished from benign seromas by CD30 ELISA and LFA, but LFA requires less time (<20 min) and can be performed without special equipment by non-specialized personnel, suggesting future point-of-care testing for BIA-ALCL may be feasible.