Efficacy of the positioning system when measuring canine intraocular pressures with the Reichert® Tono-Vera® Vet rebound tonometer.
Article in Veterinary ophthalmology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed, 2 citations in OpenAlex.
- Validation of the eyeTelemed IOPvet indentation tonometer for use in dogs.Veterinary ophthalmology · 2024Article
- Validation of the Reichert® Tono-Vera® Vet rebound tonometer in normal ex vivo canine eyes.Veterinary ophthalmology · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors at 2 institutions in 1 country.
Funding
Abstract
objectiveTo compare intraocular pressure (IOP) measurements in dogs taken with the Reichert® Tono-Vera® Vet rebound tonometer with and without the automatic positioning system. ANIMALS STUDIED: Measurements were taken on 49 eyes from 26 Beagle-derived dogs with variable genetics-four non-glaucomatous and 22 ADAMTS10-mutant dogs affected with different stages of open-angle glaucoma. Seventeen of the 26 dogs were measured 2-4 times on different days with variable intervals since IOP-lowering medications were administered. PROCEDURES: In each dog, tonometry was performed with the Tono-Vera® Vet using three different methods in a randomized order: (Method 1) Average of three readings with an automatic positioning system; (Method 2) one reading with an automatic positioning system; and (Method 3) average of three readings obtained without the automatic positioning system. Statistical analyses included one-way ANOVA, Tukey pairwise comparisons, and Bland-Altman plots (MiniTab®).
resultsWith each of the three tonometry methods, 116 measurements were taken, resulting in 348 total IOP measurements with a range of 12.8-49.9 mmHg. The means and standard deviations for each method were 25.4 ± 6.9 mmHg (Method 1), 26.0 ± 7.2 mmHg (Method 2), and 26.9 ± 7.7 mmHg (Method 3), with no significant differences (p = .27). Mean IOP variances were also not significantly different between tonometry methods (p = .24 to .78).
conclusionsBecause mean IOPs and their standard deviations were not statistically different between the three tonometry methods, we conclude that Tono-Vera® Vet measurements conducted without the aid of the positioning system still provide reliable results.