ArticleBioengineering (Basel, Switzerland)2023

In Vitro Wear of a Novel Vitamin E Crosslinked Polyethylene Lumbar Total Joint Replacement.

Ryan L Siskey, Ronald V Yarbrough, Hannah Spece, Scott D Hodges, Steven C Humphreys, Steven M Kurtz

Abstract read
In one paragraph

Article in Bioengineering (Basel, Switzerland), 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.

0numbers the graph read from it
0cells of the map it votes in
9citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

9 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Review
  7. Article
  8. Article
  9. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Ryan L SiskeyExponent Inc., Philadelphia, PA 19104, USA.
Ronald V Yarbrough3Spine, Chattanooga, TN 37402, USA.
Hannah SpeceSchool of Biomedical Engineering, Science, and Health Systems, Drexel University, Philadelphia, PA 19104, USA.
Scott D Hodges3Spine, Chattanooga, TN 37402, USA.
Steven C Humphreys3Spine, Chattanooga, TN 37402, USA.
Steven M KurtzSchool of Biomedical Engineering, Science, and Health Systems, Drexel University, Philadelphia, PA 19104, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundA novel, lumbar total joint replacement (TJR) design has been developed to treat degeneration across all three columns of the lumbar spine (anterior, middle, and posterior columns). Thus far, there has been no in vitro studies that establish the preclinical safety profile of the vitamin E-stabilized highly crosslinked polyethylene (VE-HXLPE) lumbar TJR relative to historical lumbar anterior disc replacement for the known risks of wear and impingement faced by all motion preserving designs for the lumbar spine. QUESTIONS/PURPOSE: In this study we asked, (1) what is the wear performance of the VE-HXLPE lumbar TJR under ideal, clean conditions? (2) Is the wear performance of VE-HXLPE in lumbar TJR sensitive to more aggressive, abrasive conditions? (3) How does the VE-HXLPE lumbar TJR perform under impingement conditions?

methodA lumbar TJR with bilateral VE-HXLPE superior bearings and CoCr inferior bearings was evaluated under clean, impingement, and abrasive conditions. Clean and abrasive testing were guided by ISO 18192-1 and impingement was assessed as per ASTM F3295. For abrasive testing, CoCr components were scratched to simulate in vivo abrasion. The devices were tested for 10 million cycles (MC) under clean conditions, 5 MC under abrasion, and 1 MC under impingement.

resultWear rates under clean and abrasive conditions were 1.2 ± 0.5 and 1.1 ± 0.6 mg/MC, respectively. The VE-HXLPE components demonstrated evidence of burnishing and multidirectional microscratching consistent with microabrasive conditions with the cobalt chromium spherical counterfaces. Under impingement, the wear rates ranged between 1.7 ± 1.1 (smallest size) and 3.9 ± 1.1 mg/MC (largest size). No functional or mechanical failure was observed across any of the wear modes.

conclusionsOverall, we found that that a VE-HXLPE-on-CoCr lumbar total joint replacement design met or exceeded the benchmarks established by traditional anterior disc replacements, with wear rates previously reported in the literature ranging between 1 and 15 mg/MC. CLINICAL RELEVANCE: The potential clinical benefits of this novel TJR design, which avoids long-term facet complications through facet removal with a posterior approach, were found to be balanced by the in vitro tribological performance of the VE-HXLPE bearings. Our encouraging in vitro findings have supported initiating an FDA-regulated clinical trial for the design which is currently under way.

Indexed as

abrasionantioxidantdesignhighly crosslinked polyethylene (HXLPE)impingementlumbar spinepreclinical testingtotal joint replacementvitamin Ewear

Identifiers

PMID37892928
PMCPMC10604298

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.