Evidence map›Paper›PMID 37849358›Full record

ArticleJournal of applied clinical medical physics2024

Commissioning of Mobius3D/FX for patient-specific quality assurance: The CUIMC-NewYork Presbyterian Hospital experience.

Fiona Li, Michael Price

Open access · goldAbstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
1.8field-weighted citation impact, top 16% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 5 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors at 1 institution in 1 country.

Fiona LiDepartment of Radiation Oncology, Columbia University Irving Medical Center, New York, New York, USA.
Michael PriceDepartment of Radiation Oncology, Columbia University Irving Medical Center, New York, New York, USA.
Columbia University Irving Medical Center · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeTo present the process undertaken by our institute in commissioning Mobius3D (M3D) for patient-specific quality assurance.

method168 plans were randomly selected to compare dose distribution measured with ArcCheck and dose calculated from M3D, both compared against the treatment planning system (TPS). The gamma criteria for measurement and M3D are 3%/2 mm with 10% and 50% dose thresholds, respectively. The effect of tissue inhomogeneity was investigated on 11 plans by recalculating the dose in M3D on a homogeneous phantom. Tolerance and action limits were established following the AAPM Task Group 218 recommendations.

resultsThe M3D showed high variability in gamma passing rate compared to the measurement. Twenty-three out of 168 plans had false negative dose comparisons. These plans fall under high tissue inhomogeneity like lung and metal implants, small field targets, and breast plans planned with high energy. One false negative case (0.6%) was observed. A single tolerance limit of 91% and 92% gamma passing rate for the M3D and measurement-based PSQA were established, respectively. Against the expectation, recalculating plans on the homogeneous phantom in M3D did not necessarily increase the gamma passing rate. These plans have a duty cycle >4.2, and the small field sizes combined with differences in slice thickness contributed to observed dose differences in the homogeneous phantom comparisons.

conclusionFollowing the commissioning, M3D is adopted in our institute. Currently, the gamma criteria used for measurement and M3D are 3%/2 mm, 40% dose threshold, with gamma passing rates of 92% and 95%, respectively. A higher passing rate for M3D is adopted until more data is available. The combined effect of plan modulation, the field sizes, the tissue inhomogeneity, the dose algorithm, and the volume averaging effect from differences in slice thickness can contribute to the differences in dose in M3D and TPS.

Indexed as

Radiotherapy, Intensity-ModulatedRadiotherapy Planning, Computer-AssistedAlgorithmsHumansProtestantismQuality Assurance, Health CareRadiotherapy DosageAction limitMLC Log file-based QAMobius3D/FXPatient-specific quality assuranceTolerance limit

Identifiers

PMID37849358
PMCPMC10860561
OpenAlexW4387729542

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.