Evidence map›Paper›PMID 37823742›Full record

ArticleActa orthopaedica et traumatologica turcica2023

Biomechanical comparison of fixation methods on third-generation femoral composite bone models in Pauwels type 3 femoral neck fractures: Contribution of the medial buttress plate to fixation.

Feyzi Kılıç, Abdulkadir Polat, Fatih Yamak, Ergün Bozdag, Fırat Fidan, Hakan Başar

Open access · goldAbstract read
In one paragraph

Article in Acta orthopaedica et traumatologica turcica, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
0.3field-weighted citation impact, top 37% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 1 citations in OpenAlex.

  1. Trial
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 1 institution in 1 country.

Feyzi KılıçDepartment of Orthopaedics and Traumatology, Doğubeyazıt State Hospital, Ağrı, Turkey.
Abdulkadir PolatDepartment of Orthopaedics and Traumatology, University of Health Sciences Gaziosmanpaşa Training and Research Hospital, İstanbul, Turkey.
Fatih YamakIstanbul Technical University, Faculty of Mechanical Engineering, Istanbul, Turkey.
Ergün BozdagIstanbul Technical University, Faculty of Mechanical Engineering, Istanbul, Turkey.
Fırat FidanDepartment of Orthopaedics and Traumatology, Istanbul Aydın University, Faculty of Medicine, Istanbul, Turkey.
Hakan BaşarDepartment of Orthopaedics and Traumatology, University of Health Sciences Gaziosmanpaşa Training and Research Hospital, İstanbul, Turkey.
Istanbul Technical University · TR

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectiveTo compare the use of cannulated screws in an inverted triangular configuration, medial buttress plates, and the combination of these 2 fixation methods in the treatment of Pauwels type 3 femoral neck fractures.

methodsTwenty-eight anatomical composite third-generation femoral bone models were divided into 4 groups. The control group (group 1) was formed with 7 third-generation intact bone models. The fracture model was created with a 70° cutting block to fit 21 Pauwels type 3 fracture configurations. Seven models were fixed with an isolated 3.5 mm one-third semi-tubular medial buttress plate (group 2), 7 were fixed in an inverted triangular configuration with 6.5 mm cannulated screws (group 3), and 7 were fixed using a combination of 6.5 mm cannulated screws and a medial buttress plate (group 4). Cyclic loading was applied using axial forces ranging from 60 N to 600 N and moments ranging from 0.7 Nm to 7.0 Nm for 500 cycles. Once the cyclic loading stage was completed, the loads were removed from the system, and the quasi-static loading stage was employed to determine the stiffness and failure forces of the system under both axial and torsional forces. Quasi-static tests were performed with an axial speed of 1.8 mm/min and a torsional speed of 4.5°/min. The biomechanical properties of all groups were examined in terms of axial stiffness, torsional stiffness, and maximum axial force parameters.

resultsThe stiffness values of groups 1, 2, 3, and 4 were 303 ± 35.8 N/mm, 159.6 ± 25 N/mm, 232 ± 35.9 N/mm, and 366.9 ± 58 N/mm, respectively, under axial forces (P < .01); 2172.7 ± 252.1 Nmm/°, 1225.3 ± 238.6 Nmm/°, 2123 ± 359.4 Nmm/°, and 2721.85 ± 304 Nmm/°, respectively, under torsional moments (P < .01); and 2072.1 ± 256.1 N, 1379.9 ± 290.6 N, 2099.1 ± 454.2 N, and 2648.4 ± 364.6 N, respectively, under the maximum force (P < .01).

conclusionThis study showed that in the fixation of Pauwels type 3 fractures formed on third-generation bone models, the utilization of half-thread cannulated screws in an inverted triangle configuration, along with a medial buttress plate, provided stronger fixation compared to the remaining implant groups and the control group. According to the evaluation of the parameters, the isolated application of a medial buttress plate had poorer biomechanical properties than other fixation methods.

Indexed as

Bone ScrewsFemoral Neck FracturesBiomechanical PhenomenaBone PlatesFemurFracture Fixation, InternalHumans