Escape of surgical smoke particles, comparing conventional and valveless trocar systems.

Daniel Robertson, Frank Sterke, Willem van Weteringen, Alberto Arezzo, Yoav Mintz, Technology Committee of the European Association for Endoscopic Surgery (EAES), Tim Horeman

Open access · hybridAbstract read
In one paragraph

Article in Surgical endoscopy, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
0.5field-weighted citation impact, top 38% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 3 citations in OpenAlex.

  1. Article
  2. Review
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 5 institutions in 3 countries.

Daniel Robertson *Department of Biomechanical Engineering, Faculty of Mechanical Engineering, Delft University of Technology, Mekelweg 2, 2628 CD, Delft, The Netherlands. p.d.robertson@tudelft.nl.ORCID 0000-0002-3373-5113
Frank Sterke *Department of Biomechanical Engineering, Faculty of Mechanical Engineering, Delft University of Technology, Mekelweg 2, 2628 CD, Delft, The Netherlands.
Willem van WeteringenDepartment of Pediatric Surgery, Erasmus MC Sophia Children's Hospital, University Medical Center Rotterdam, Rotterdam, The Netherlands.
Alberto ArezzoDepartment of Surgical Sciences, University of Torino, Turin, Italy.
Yoav MintzDepartment of General Surgery, Hadassah Hebrew University Medical Center, Jerusalem, Israel.
Technology Committee of the European Association for Endoscopic Surgery (EAES)
Tim HoremanDepartment of Biomechanical Engineering, Faculty of Mechanical Engineering, Delft University of Technology, Mekelweg 2, 2628 CD, Delft, The Netherlands.
Delft University of Technology · NLErasmus MC · NLErasmus MC - Sophia Children’s Hospital · NLHadassah Medical Center · ILUniversity of Turin · IT

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundDuring minimal access surgery, surgical smoke is produced which can potentially be inhaled by the surgical team, leading to several health risks. This smoke can escape from the abdominal cavity into the operating room due to trocar leakage. The trocars and insufflator that are used during surgery influence gas leakage. Therefore, this study compares particle escape from a valveless (Conmed AirSeal iFS), and a conventional (Karl Storz Endoflator) system. MATERIALS AND

methodsUsing an in vitro model, a conventional and a valveless trocar system were compared. A protocol that simulated various surgical phases was defined to assess the surgical conditions and particle leakage. Insufflation pressures and instrument diameters were varied as these are known to affect gas leakage.

resultsThe conventional trocar leaked during two distinct phases. Removal of the obturator caused a sudden release of particles. During instrument insertion, an average of 211 (IQR 111) particles per second escaped when using the 5 mm diameter instrument. With the 10 mm instrument, 50 (IQR 13) particles per second were measured. With the conventional trocar, a higher abdominal pressure increased particle leakage. The valveless trocar demonstrated a continuously high particle release during all phases. After the obturator was removed, particle escape increased sharply. Particle escape decreased to 1276 (IQR 580) particles per second for the 5 mm instrument insertion, and 1084 (IQR 630) particles per second for 10 mm instrument insertion. With the valveless trocar system, a higher insufflation pressure lowered particle escape.

conclusionsThis study shows that a valveless trocar system releases more particles into the operating room environment than a conventional trocar. During instrument insertion, the leakage through the valveless system is 6 to 20 times higher than the conventional system. With a valveless trocar, leakage decreases with increasing pressure. With both trocar types leakage depends on instrument diameter.

Indexed as

Abdominal CavityInsufflationLaparoscopyHumansSmokeSurgical InstrumentsSmokeInsufflation pressureLaparoscopySurgical smokeValveless insufflation