Evidence map›Paper›PMID 37792776›Full record

ArticlePloS one2023

The effect of 3D stereopsis and hand-tool alignment on learning effectiveness and skill transfer of a VR-based simulator for dental training.

Maximilian Kaluschke, Myat Su Yin, Peter Haddawy, Siriwan Suebnukarn, Gabriel Zachmann

Open access · goldAbstract read
In one paragraph

Article in PloS one, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed, 1 pooled it
6.9field-weighted citation impact, top 3% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed, 1 synthesis or guideline pooled it, 14 citations in OpenAlex.

  1. Pooled it
  2. Trial
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 3 institutions in 2 countries.

Maximilian KaluschkeComputer Graphics and Virtual Reality, University of Bremen, Bremen, Germany.ORCID 0000-0001-5797-6344
Myat Su YinFaculty of ICT, Mahidol University, Bangkok, Thailand.ORCID 0000-0001-6483-7393
Peter HaddawyFaculty of ICT, Mahidol University, Bangkok, Thailand.ORCID 0000-0003-2203-006X
Siriwan SuebnukarnFaculty of Dentistry, Thammasat University, Bangkok, Thailand.ORCID 0000-0003-1237-1274
Gabriel ZachmannComputer Graphics and Virtual Reality, University of Bremen, Bremen, Germany.
Mahidol University · THUniversity of Bremen · DEThammasat University · TH

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Recent years have seen the proliferation of VR-based dental simulators using a wide variety of different VR configurations with varying degrees of realism. Important aspects distinguishing VR hardware configurations are 3D stereoscopic rendering and visual alignment of the user's hands with the virtual tools. New dental simulators are often evaluated without analysing the impact of these simulation aspects. In this paper, we seek to determine the impact of 3D stereoscopic rendering and of hand-tool alignment on the teaching effectiveness and skill assessment accuracy of a VR dental simulator. We developed a bimanual simulator using an HMD and two haptic devices that provides an immersive environment with both 3D stereoscopic rendering and hand-tool alignment. We then independently controlled for each of the two aspects of the simulation. We trained four groups of students in root canal access opening using the simulator and measured the virtual and real learning gains. We quantified the real learning gains by pre- and post-testing using realistic plastic teeth and the virtual learning gains by scoring the training outcomes inside the simulator. We developed a scoring metric to automatically score the training outcomes that strongly correlates with experts' scoring of those outcomes. We found that hand-tool alignment has a positive impact on virtual and real learning gains, and improves the accuracy of skill assessment. We found that stereoscopic 3D had a negative impact on virtual and real learning gains, however it improves the accuracy of skill assessment. This finding is counter-intuitive, and we found eye-tooth distance to be a confounding variable of stereoscopic 3D, as it was significantly lower for the monoscopic 3D condition and negatively correlates with real learning gain. The results of our study provide valuable information for the future design of dental simulators, as well as simulators for other high-precision psycho-motor tasks.

Indexed as

Computer-Assisted InstructionVirtual RealityClinical CompetenceComputer SimulationHumansLearningUser-Computer Interface

Identifiers

PMID37792776
PMCPMC10550175
OpenAlexW4387331466

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked