ArticleArchives of orthopaedic and trauma surgery2024

AI-based hip prosthesis failure prediction through evolutional radiological indices.

Matteo Bulloni, Francesco Manlio Gambaro, Katia Chiappetta, Guido Grappiolo, Valentina Corino, Mattia Loppini

Abstract read
PubMed Publisher
In one paragraph

Article in Archives of orthopaedic and trauma surgery, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed.

  1. Review
  2. Review
  3. Article
  4. Review
  5. Article
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Matteo BulloniDepartment of Electronics, Information and Bioengineering, Politecnico di Milano, Via Ponzio 34/5, 20133, Milan, Italy.
Francesco Manlio GambaroDepartment of Biomedical Sciences, Humanitas University, Via Rita Levi Montalcini 4, 20090, Pieve Emanuele, MI, Italy.
Katia ChiappettaIRCCS Humanitas Research Hospital, Via Manzoni 56, 20089, Rozzano, MI, Italy.
Guido GrappioloIRCCS Humanitas Research Hospital, Via Manzoni 56, 20089, Rozzano, MI, Italy.
Valentina CorinoDepartment of Electronics, Information and Bioengineering, Politecnico di Milano, Via Ponzio 34/5, 20133, Milan, Italy.
Mattia LoppiniDepartment of Biomedical Sciences, Humanitas University, Via Rita Levi Montalcini 4, 20090, Pieve Emanuele, MI, Italy. mattia.loppini@hunimed.eu.ORCID http://orcid.org/0000-0001-8569-0669

Funding

Ministero della Salute GR-2018-12367275
6 · The paper itself

Abstract

introductionThis study aimed to develop artificial intelligence models for predicting hip implant failure from radiological features. Analyzing the evolution of the periprosthetic bone and implant's position throughout the entire follow-up period has shown the potential to be more relevant in outcome prediction than simply considering the latest radiographic images. Thus, we investigated an AI-based model employing a small set of evolutional parameters derived from conventional radiological features to predict hip prosthesis failure. MATERIALS AND

methodsOne hundred sixty-nine radiological features were annotated from historical anteroposterior and lateral radiographs for 162 total hip arthroplasty patients, 32 of which later underwent implant failure. Linear regression on each patient's chronologically sorted radiological features was employed to derive 169 corresponding evolutional parameters per image. Three sets of machine learning predictors were developed: one employing the original features (standard model), one the evolutional ones (evolutional model), and the last their union (hybrid model). Each set included a model employing all the available features (full model) and a model employing the few most predictive ones according to Gini importance (minimal model).

resultsThe evolutional and hybrid predictors resulted highly effective (area under the ROC curve (AUC) of full models = 0.94), outperforming the standard one, whose AUC was only 0.82. The minimal hybrid model, employing just four features, three of which evolutional, scored an AUC of 0.95, proving even more accurate than the full one, exploiting 173 features. This tool could be shaped to be either highly specific (sensitivity: 80%, specificity: 98.6%) or highly sensitive (sensitivity: 90%, specificity: 92.4%).

conclusionThe proposed predictor may represent a highly sensitive screening tool for clinicians, capable to predict THA failure with an advance between a few months and more than a year through only four radiological parameters, considering either their value at the latest visit or their evolution through time.

Indexed as

Arthroplasty, Replacement, HipHip ProsthesisArtificial IntelligenceHumansProsthesis FailureRadiographyRetrospective StudiesAIHipRadiographTHA