Trial reportOrthopaedic surgery2023
Comparison between Three-Dimensional Printed Titanium and PEEK Cages for Cervical and Lumbar Interbody Fusion: A Prospective Controlled Trial.
Trial report in Orthopaedic surgery, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 18 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
18 citing papers in PubMed, 24 citations in OpenAlex.
- Anterior cervical discectomy and fusion without anchoring: a multicenter comparison of stand-alone porous titanium versus anchored PEEK interbody cages.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2026Article
- Outcomes of 3D-printed titanium cages for surgical reconstruction in infectious spondylodiscitis: a systematic review.Journal of spine surgery (Hong Kong) · 2026Review
- Material and Geometric Innovations in Lumbar and Cervical Interbody Devices: A Systematic Scoping Review.Annals of biomedical engineering · 2026Review
- C- VBQ versus CT Hounsfield units in predicting cage subsidence following ACDF: a retrospective cohort analysis.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2026Article
- Optimal cage biomaterials for fusion and subsidence outcomes in lumbar spinal fusion: a network meta-analysis of 4881 patients.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2026Review
- Clinical outcomes and radiologic parameters of percutaneous endoscopic posterior/transforaminal lumbar interbody fusion using rammed-earth technique.Frontiers in surgery · 2026Article
- Is Ti-Coated PEEK Superior to PEEK for Lumbar and Cervical Fusion Procedures? A Systematic Review and Meta-Analysis.Journal of clinical medicine · 2025Review
- Evaluating the Efficacy of a Novel Titanium Cage System in ALIF and LLIF: A Retrospective Clinical and Radiographic Analysis.Journal of clinical medicine · 2025Article
- Comparison of osseointegration in 3D-printed hexanium titanium cages and PEEK cervical cages: In-vivo ovine cervical fusion model.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2025Article
- Article
- Evaluation of Healthcare Outcomes of Patients Treated with 3D-Printed-Titanium and PEEK Cages During Fusion Procedures in the Lumbar Spine.Medical devices (Auckland, N.Z.) · 2025Article
- Personalized porous tantalum implants crafted via 3D printing: new horizons in complex cervical-thoracic spinal fusion.Frontiers in bioengineering and biotechnology · 2025Article
- Innovative 3D printing technologies and advanced materials revolutionizing orthopedic surgery: current applications and future directions.Frontiers in bioengineering and biotechnology · 2025Review
- Top Three Healthcare Trends in Orthopaedics Affecting Musculoskeletal Care in the Elderly.Current aging science · 2025Review
- Advancements in Custom 3D-Printed Titanium Interbody Spinal Fusion Cages and Their Relevance in Personalized Spine Care.Journal of personalized medicine · 2024Article
- Expandable Cages for Lumbar Interbody Fusion: A Narrative Review.Journal of clinical medicine · 2024Review
- 3D-printed titanium alloy cage in anterior and lateral lumbar interbody fusion for degenerative lumbar spine disease.Journal of spine surgery (Hong Kong) · 2024Article
- An analysis of different modalities of bone mineral densitometry evaluation in cage subsidence in anterior cervical discectomy and fusion.Frontiers in surgery · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors at 1 institution in 2 countries.
Funding
Abstract
objectivesThe three-dimensional printing titanium (3DPT) cage with excellent biomechanical properties and osseointegration capabilities has been initially used in spinal fusion, while the polyetheretherketone (PEEK) cage, a bioinert material device, has been a widely used for decades with relatively excellent clinical outcomes. This study was performed to investigate the early radiographic and clinical outcomes of 3DPT cage versus PEEK cage in patients undergoing anterior cervical discectomy and fusion (ACDF) and transforaminal lumbar interbody fusion (TLIF).
methodsThis prospective controlled trial, from December 2019 to June 2022, included patients undergoing ACDF and TLIF with 3DPT cages and compared them to patients using PEEK cages for treating spinal degenerative disorders. The outcome measures included radiographic parameters (intervertebral height [IH], subsidence, fusion status, and bone-cage interface contact) and clinical outcomes (Japanese Orthopaedic Association [JOA], Neck Disability Index [NDI], Oswestry Disability Index [ODI], Short Form 12-Item Survey [SF-12], Visual Analog Scale [VAS], and Odom's criteria). Student's independent samples t test and Pearson's chi-square test were used to compare the outcome measures between the two groups before surgery and at 1 week, 3 and 6 months after surgery.
resultsFor the patients undergoing ACDF, the 3DPT (18 patients/[26 segments]) and PEEK groups (18 patients/[26 segments]) had similar fusion rates at 3 months and 6 months follow-up (3 months: 96.2% vs. 83.3%, p = 0.182; 6 months: 100% vs. 91.7%, p = 0.225). The subsidence in the 3DPT group was significantly lower than that in the PEEK group (3 months: 0.4 ± 0.2 mm vs. 0.9 ± 0.7 mm p = 0.004; 6 months: 0.7 ± 0.3 mm vs. 1.5 ± 0.8 mm, p < 0.001). 3DPT and PEEK cage all achieved sufficient contact with the cervical endplates. For the patients undergoing TLIF, the 3DPT (20 patients/[26 segments]) and PEEK groups (20 patients/[24 segments]) had no statistical difference in fusion rate (3 months: 84.6% vs. 58.3%, p = 0.059; 6 months: 92.3% vs. 75%, p = 0.132). The subsidence was lower than that in the PEEK group without significantly difference (3 months: 0.9 ± 0.7 mm vs.1.2 ± 0.9 mm p = 0.136; 6 months: 1.6 ± 1.0 mm vs. 2.0 ± 1.0 mm, p = 0.200). At the 3-month follow-up, the bone-cage interface contact of the 3DPT cage was significantly better than that of the PEEK cage (poor contact: 15.4% vs. 75%, p < 0.001). The values of UAR were higher in the 3DPT group than in the PEEK group during the follow-up in cervical and lumbar fusion, there were more statistical differences in lumbar fusion. There were no significant differences in the clinical assessment between 3DPT or PEEK cage in spinal fusion.
conclusionThe 3DPT cage and PEEK cage can achieve excellent clinical outcomes in cervical and lumbar fusion. The 3DPT cage has advantage in fusion quality, subsidence severity, and bone-cage interface contact than PEEK cage.