Evaluation of the Humidification Performances of New Generation of Heated Wire Humidifiers.
Article in Respiratory care, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Consequences of Pausing Heated Humidification During Invasive Ventilation.Respiratory care · 2024Article
- In-Line Aerosol Therapy via Nasal Cannula during Adult and Paediatric Normal, Obstructive, and Restrictive Breathing.Pharmaceutics · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundAt high ambient temperatures in ICU rooms, the humidification performances of heated-wire humidifiers are significantly reduced, with delivered gas humidity well below 30 mg H
methodsWe measured, on the bench, the hygrometry of inspiratory gases delivered by a new generation of heated-wire humidifiers (i) FP950, (ii) VHB20 and a previous generation of heated-wire humidifiers, (iii) MR850 (Fisher & Paykel) with the usual settings (37°C at the chamber/40°C at the Y-piece), (iv) MR850 with no temperature gradient (40°C/40°C), and (v) MR850 with the automatic compensation algorithm activated. Hygrometry was measured with the psychrometric method after 1 h of stability while varying the room temperature from 20 to 30°C.
resultsTwo hundred ninety-four hygrometric bench measurements were performed at steady state for the different tested conditions. With the new heated-wire humidifiers (FP950 and VHB20), gas humidity delivered remained > 30 mg H
conclusionsThe new FP950 and VHB20 heated-wire humidifiers by using advanced algorithms demonstrated stable performance while varying the ambient temperature by 20-30°C, better than the previous generation of heated humidifiers when ambient temperatures were high.