Evidence map›Paper›PMID 37689499›Full record

ArticleZeitschrift fur medizinische Physik2025

Generation, validation, and benchmarking of a commercial independent Monte Carlo calculation beam model for multi-target SRS.

Justus Adamson, Brett G Erickson, Chunhao Wang, Yunfeng Cui, Markus Alber, John Kirkpatrick, Fang-Fang Yin

Open access · goldAbstract readValidation Study
In one paragraph

Article in Zeitschrift fur medizinische Physik, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
2.9field-weighted citation impact, top 9% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 8 citations in OpenAlex.

  1. Article
  2. Review
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 2 institutions in 1 country.

Justus AdamsonDuke University Medical Center, Department of Radiation Oncology, Durham, NC, USA. Electronic address: justus.adamson@duke.edu.
Brett G EricksonDuke University Medical Center, Department of Radiation Oncology, Durham, NC, USA.
Chunhao WangDuke University Medical Center, Department of Radiation Oncology, Durham, NC, USA.
Yunfeng CuiDuke University Medical Center, Department of Radiation Oncology, Durham, NC, USA.
Markus AlberScientific RT GmbH, Munich, Germany.
John KirkpatrickDuke University Medical Center, Department of Radiation Oncology, Durham, NC, USA.
Fang-Fang YinDuke University Medical Center, Department of Radiation Oncology, Durham, NC, USA.
Duke Medical Center · USDuke University · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundDosimetric validation of single isocenter multi-target radiosurgery plans is difficult due to conditions of electronic disequilibrium and the simultaneous irradiation of multiple off-axis lesions dispersed throughout the volume. Here we report the benchmarking of a customizable Monte Carlo secondary dose calculation algorithm specific for multi-target radiosurgery which future users may use to guide their commissioning and clinical implementation. PURPOSE: To report the generation, validation, and clinical benchmarking of a volumetric Monte Carlo (MC) dose calculation beam model for single isocenter radiosurgery of intracranial multi-focal disease.

methodsThe beam model was prepared within SciMoCa (ScientificRT, Munich Germany), a commercial independent dose calculation software, with the aim of broad availability via the commercial software for use with single isocenter radiosurgery. The process included (1) definition & acquisition of measurement data required for beam modeling, (2) tuning model parameters to match measurements, (3) validation of the beam model via independent measurements and end-to-end testing, and finally, (4) clinical benchmarking and validation of beam model utility in a patient specific QA setting. We utilized a 6X Flattening-Filter-Free photon beam from a TrueBeam STX linear accelerator (Siemens Healthineers, Munich Germany).

resultsIn addition to the measured data required for standard IMRT/VMAT (depth dose, central axis profiles & output factors, leaf gap), beam modeling and validation for single-isocenter SRS required central axis and off axis (5 cm & 9 cm) small field output factors and comparison between measurement and simulation of backscatter with aperture for jaw much greater than MLCs. Validation end-to-end measurements included SRS MapCHECK in StereoPHAN geometry (2%/1 mm Gamma = 99.2% ± 2.2%), and OSL & scintillator measurements in anthropomorphic STEEV phantom (6 targets, volume = 0.1-4.1cc, distance from isocenter = 1.2-7.9 cm) for which mean difference was -1.9% ± 2.2%. For 10 patient cases, MC for individual PTVs was -0.8% ± 1.5%, -1.3% ± 1.7%, and -0.5% ± 1.8% for mean dose, D

conclusionsThe beam modeling, validation, and clinical action criteria outlined here serves as a benchmark for future users of the customized beam model within SciMoCa for single isocenter radiosurgery of multi-focal disease.

Indexed as

Monte Carlo MethodRadiosurgeryRadiotherapy Planning, Computer-AssistedAlgorithmsBenchmarkingHumansPhantoms, ImagingRadiometryRadiotherapy DosageMonte CarloRadiosurgeryTG-219

Identifiers

PMID37689499
PMCPMC12664461
OpenAlexW4386582467

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.