ArticleJournal of applied clinical medical physics2023
Preliminary study on dosimetry characteristics of a novel cylindrical dose verification system.
Article in Journal of applied clinical medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 2 citations in OpenAlex.
- Preliminary study on dosimetry characteristics of a novel cylindrical dose verification system.Journal of applied clinical medical physics · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors at 4 institutions in 1 country.
Funding
Abstract
objectiveTo develop a novel ionization chamber array dosimetry system, study its dosimetry characteristics, and perform preliminary tests for plan dose verification.
methodsThe dosimetry characteristics of this new array were tested, including short-term and long-term reproducibility, dose linearity, dose rate dependence, field size dependence, and angular dependence. The open field and MLC field plans were designed for dose testing. Randomly select 30 patient treatment plans (10 intensity-modulated radiation therapy [IMRT] plans and 20 volumetric modulated arc therapy [VMAT] plans) that have undergone dose verification using Portal Dosimetry to perform verification measurement and evaluate dose verification test results.
resultsThe dosimetry characteristics of the arrays all performed well. The gamma passing rates (3%/2 mm) were more than 96% for the combined open field and MLC field plans. The average gamma pass rates were (99.54 ± 0.58)% and (96.70 ± 3.41)% for the 10 IMRT plans and (99.32 ± 0.89)% and (94.91 ± 6.01)% for the 20 VMAT plans at the 3%/2 mm and 2%/2 mm criteria, respectively, which is similar to the Portal Dosimetry's measurement results.
conclusionsThis novel ionization chamber array demonstrates good dosimetry characteristics and is suitable for clinical IMRT and VMAT plan verifications.