Novel T1 Analysis Method to Address Reduced Measurement Accuracy Due to Irregular Heart Rate Variability in Myocardial T1 Mapping Using Polarity-corrected Inversion Time Preparation.
Article in Magnetic resonance in medical sciences : MRMS : an official journal of Japan Society of Magnetic Resonance in Medicine, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposePolarity-corrected inversion time preparation (PCTIP), a myocardial T1 mapping technique, is expected to reduce measurement underestimation in the modified Look-Locker inversion recover method. However, measurement precision is reduced, especially for heart rate variability. We devised an analysis using a recurrence formula to overcome this problem and showed that it improved the measurement accuracy, especially at high heart rates. Therefore, this study aimed to determine the effect of this analysis on the accuracy and precision of T1 measurements for irregular heart rate variability.
methodsA PCTIP scan using a 3T MRI scanner was performed in phantom experiment. We generated the simulated R-waves required for electrocardiogram (ECG)-gated acquisition using a signal generator set to 30 combinations. T1 map was generated using the signal train of the PCTIP images by nonlinear curve fitting using conventional and recurrence formulas. Accuracy against reference T1 and precision of heart rate variability were evaluated. To evaluate the fitting accuracy of both analyses, the relative fitting error was calculated.
resultsFor the longer T1, the fitting error was larger than the short T1, with the conventional analysis showing 10.1±2.0%. The recurrence formula analysis showed a small fitting error less than 1%, which was consistent for all heart rate variability patterns. In the conventional analysis, the accuracy, especially for longer T1, showed a large underestimation of the measurements and poor linearity. However, in the recurrence formula analysis, the accuracy improved at a long T1, and linearity also improved. The Bland-Altman plot showed that it varied greatly depending on the heart rate variability pattern for the longer T1 in the conventional analysis, whereas the recurrence formula analysis suppressed this variation.
conclusionT1 analysis of PCTIP using the recurrence formula analysis achieved accurate and precise T1 measurements, even for irregular heart rate variability.