Observational studyEuropace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology2023
Performance of a multi-sensor implantable defibrillator algorithm for heart failure monitoring in the presence of atrial fibrillation.
Observational study in Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT02275637 (Arrhythmias Detection in a Real World Population), which is not on this map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Arrhythmias Detection in a Real World Population: the RHYTHM DETECT Registry
Who cites it
2 citing papers in PubMed, 7 citations in OpenAlex.
- Association between amount of biventricular pacing and heart failure status measured by a multisensor implantable defibrillator algorithm.Cardiovascular digital health journal · 2024Article
- Predicting worsening heart failure hospitalizations in patients with implantable cardioverter defibrillators: is it all about alerts? A pooled analysis of nine trials.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
20 authors at 15 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
aimsThe HeartLogic Index combines data from multiple implantable cardioverter defibrillators (ICDs) sensors and has been shown to accurately stratify patients at risk of heart failure (HF) events. We evaluated and compared the performance of this algorithm during sinus rhythm and during long-lasting atrial fibrillation (AF). METHODS AND
resultsHeartLogic was activated in 568 ICD patients from 26 centres. We found periods of ≥30 consecutive days with an atrial high-rate episode (AHRE) burden <1 h/day and periods with an AHRE burden ≥20 h/day. We then identified patients who met both criteria during the follow-up (AHRE group, n = 53), to allow pairwise comparison of periods. For control purposes, we identified patients with an AHRE burden <1 h throughout their follow-up and implemented 2:1 propensity score matching vs. the AHRE group (matched non-AHRE group, n = 106). In the AHRE group, the rate of alerts was 1.2 [95% confidence interval (CI): 1.0-1.5]/patient-year during periods with an AHRE burden <1 h/day and 2.0 (95% CI: 1.5-2.6)/patient-year during periods with an AHRE-burden ≥20 h/day (P = 0.004). The rate of HF hospitalizations was 0.34 (95% CI: 0.15-0.69)/patient-year during IN-alert periods and 0.06 (95% CI: 0.02-0.14)/patient-year during OUT-of-alert periods (P < 0.001). The IN/OUT-of-alert state incidence rate ratio of HF hospitalizations was 8.59 (95% CI: 1.67-55.31) during periods with an AHRE burden <1 h/day and 2.70 (95% CI: 1.01-28.33) during periods with an AHRE burden ≥20 h/day. In the matched non-AHRE group, the rate of HF hospitalizations was 0.29 (95% CI: 0.12-0.60)/patient-year during IN-alert periods and 0.04 (95% CI: 0.02-0.08)/patient-year during OUT-of-alert periods (P < 0.001). The incidence rate ratio was 7.11 (95% CI: 2.19-22.44).
conclusionPatients received more alerts during periods of AF. The ability of the algorithm to identify increased risk of HF events was confirmed during AF, despite a lower IN/OUT-of-alert incidence rate ratio in comparison with non-AF periods and non-AF patients. CLINICAL
trial registrationhttp://clinicaltrials.gov/Identifier: NCT02275637.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.