ArticleIEEE transactions on bio-medical engineering2024
A Novel Vascular Anastomotic Coupling Device for End-to-End Anastomosis of Arteries and Veins.
Article in IEEE transactions on bio-medical engineering, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed, 2 citations in OpenAlex.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors at 1 institution in 1 country.
Funding
Abstract
objectiveHand-sutured (HS) techniques remain the gold standard for most microvascular anastomoses in microsurgery. HS techniques can result in endothelial lacerations and back wall suturing, leading to complications such as thrombosis and free tissue loss. A novel force-interference-fit vascular coupling device (FIF-VCD) system can potentially reduce the need for HS and improve end-to-end anastomosis. This study aims to describe the development and testing of a novel FIF-VCD system for 1.5 to 4.0 mm outside diameter arteries and veins.
methodsBenchtop anastomoses were performed using porcine cadaver arteries and veins. Decoupling force and anastomotic leakage were tested under simulated worst-case intravital physiological conditions. The 1.5 mm FIF-VCD system was used to perform cadaver rat abdominal aorta anastomoses.
resultsBenchtop testing showed that the vessels coupled with the FIF-VCD system could withstand simulated worst-case intravital physiological conditions with a 95% confidence interval for the average decoupling force safety factor of 8.2 ± 1.0 (5.2 ± 1.0 N) and a 95% confidence interval for the average leakage rate safety factor of 26 ± 3.6 (8.4 ± 0.14 and 95 ± 1.4 μL/s at 150 and 360 mmHg, respectively) when compared to HS anastomotic leakage rates (310 ± 14 and 2,100 ± 72 μL/s at 150 and 360 mmHg, respectively). The FIF-VCD system was successful in performing cadaver rat abdominal aorta anastomoses.
conclusionThe FIF-VCD system can potentially replace HS in microsurgery, allowing the safe and effective connection of arteries and veins. Further studies are needed to confirm the clinical viability and effectiveness of the FIF-VCD system.