Evidence map›Paper›PMID 37608240›Full record

ArticleClinical oral investigations2023

Biomechanical behavior of implants with different diameters in relation to simulated bone loss- an in vitro study.

Tobias Graf, Jan-Frederik Güth, Josef Schweiger, Kurt-Jürgen Erdelt, Daniel Edelhoff, Michael Stimmelmayr

Abstract read
In one paragraph

Article in Clinical oral investigations, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Article
  2. Review
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Tobias GrafDepartment of Prosthetic Dentistry, Center for Dentistry and Oral Medicine, Goethe University Frankfurt, Theodor-Stern-Kai 7, 60596, Frankfurt am Main, Germany. T.Graf@med.uni-frankfurt.de.ORCID http://orcid.org/0000-0002-2391-5399
Jan-Frederik GüthDepartment of Prosthetic Dentistry, Center for Dentistry and Oral Medicine, Goethe University Frankfurt, Theodor-Stern-Kai 7, 60596, Frankfurt am Main, Germany.
Josef SchweigerDepartment of Prosthetic Dentistry, University Hospital, Ludwig-Maximilians-University Munich, Goethestraße 70, 80336, Munich, Germany.
Kurt-Jürgen ErdeltDepartment of Prosthetic Dentistry, University Hospital, Ludwig-Maximilians-University Munich, Goethestraße 70, 80336, Munich, Germany.
Daniel EdelhoffDepartment of Prosthetic Dentistry, University Hospital, Ludwig-Maximilians-University Munich, Goethestraße 70, 80336, Munich, Germany.
Michael StimmelmayrDepartment of Prosthetic Dentistry, University Hospital, Ludwig-Maximilians-University Munich, Goethestraße 70, 80336, Munich, Germany.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectivesBone resorption around implants could influence the resistance of the implant abutment complex (IAC). The present in vitro study aimed to assess the stability to static fatigue of implants presenting different levels of bone losses and diameters. MATERIALS AND

methodsNinety implants with an internal conical connection with 3 different implant diameters (3.3 mm (I33), 3.8 mm (I38), and 4.3 mm (I43)) and 3 simulated bone loss settings (1.5 mm (I_15), 3.0 mm (I_30), and 4.5 mm (I_45) (n = 10)) were embedded and standard abutments were mounted. All specimens were artificially aged (1,200,000 cycles, 50 N, simultaneous thermocycling) and underwent subsequently load-to-fracture test. For statistical analysis, Kolmogorov-Smirnov test, Kruskal-Wallis test, and Mann-Whitney U test (p < 0.05) were applied.

resultsAll test specimens withstood the artificial aging without damage. The mean failure values were 382.1 (± 59.2) N (I3315), 347.0 (± 35.7) N (I3330), 315.9 N (± 30.9) (I3345), 531.4 (± 36.2) N (I3815), 514.5 (± 40.8) N (I3830), 477.9 (± 26.3) N (I3845), 710.1 (± 38.2) N (I4315), 697.9 (± 65.2) N (I4330), and 662.2 N (± 45.9) (I4345). The stability of the IACs decreased in all groups when bone loss inclined. Merely, the failure load values did not significantly differ among subgroups of I43.

conclusionsLarger implant diameters and minor circular bone loss around the implant lead to a higher stability of the IAC. The smaller the implant diameter was, the more the stability was affected by the circumferential bone level. CLINICAL RELEVANCE: Preserving crestal bone level is important to ensure biomechanical sustainability at implant systems with a conical interface. It seems sensible to take the effect of eventual bone loss around implants into account during implant planning processes and restorative considerations.

Indexed as

Bone ResorptionDental ImplantsDental AbutmentsDental Implant-Abutment DesignDental Stress AnalysisHumansTitaniumDental ImplantsTitaniumBone levelDental implantsFatigue testingImplant-abutment interfaceImplant diameterStability

Identifiers

PMID37608240
PMCPMC10560161

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked