Factors affecting insulation failure in reusable surgical devices.

Takahiro Homma, Hirofumi Uehara, Hisashi Saji

Abstract read
In one paragraph

Article in Scientific reports, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Takahiro HommaDivision of Thoracic Surgery, Kurobe City Hospital, Toyama, Japan. hommat.tym@gmail.com.
Hirofumi UeharaCentral Sterile Supply Department, Toyama University Hospital, Toyama, Japan.
Hisashi SajiDepartment of Chest Surgery, St. Marianna University School of Medicine, 2-16-1 Sugao, Kawasaki, Kanagawa, 216-8511, Japan.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

The purpose of this study was to investigate the associated factors of insulation failure (IF) in reusable endoscopic instruments. The insulation coating of reusable endoscopic instruments underwent routine visual checks, hand washing to remove visible stains, and mechanized sterilization. We recorded the cleaning number and usage period of all instruments. The instruments were tested for IF using a detector. IF was found in eight of 69 devices (11.6%). Examining by clinical specialty, we found IF in 4 of 28 gastrointestinal (14.3%), 3 of 20 gynecological (15.0%), 1 of 12 urological (8.3%), and none of the nine thoracic devices. The median distance from the tip to the damaged part was 5 cm (3-5 cm). In the IF and the intact groups, the period of use [7 years (6-8) versus 7 years (4-8), P = 0.90] and the number of cleanings [281 (261-323) versus 261 (179-320), P = 0.27] were not significantly different. The IF group included products of three different companies; however, six of the eight (75.0%) were from the same company. Cleaning methods and usage period have a lower impact on IF. The use of reusable forceps as a monopolar device was found to pose a higher risk, requiring regular assessments.

Indexed as

Coloring AgentsRecordsStaining and LabelingSterilizationSterilization, ReproductiveColoring Agents

Identifiers

PMID37608091
PMCPMC10444886

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.