ArticleCardiovascular engineering and technology2023
Transcatheter Pulmonary Artery Banding in High-Risk Neonates: In-Vitro Study Provoked by Initial Clinical Experience.
Article in Cardiovascular engineering and technology, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed, 1 synthesis or guideline pooled it.
- Innovations in transcatheter interventions for hypoplastic left heart syndrome and borderline left heart structures from fetal to neonatal life: current evidence and systematic review of recent advances (2016-2025).European journal of pediatrics · 2026Pooled it
- Immediate Hemodynamic Effects of Pulmonary Artery Flow Restrictors in Univentricular Circulation: Hypothesis-Generating Insights from High-Fidelity Physiologic Streaming Data Using Machine-Learning.Pediatric cardiology · 2026Article
- Percutaneous Pulmonary Artery Banding and Duct Occlusion in a Low-Birth-Weight Preterm Infant.JACC. Case reports · 2026Article
- Fenestrated PECA exGraft Patch as a Novel Modifiable Internal Pulmonary Artery Band: A Case Report.Pediatric cardiology · 2026Article
- Proposal: Bold New Indications for Transcatheter Pulmonary Flow Restrictors.Pediatric cardiology · 2026Review
- Duct Stenting in Duct-Dependent Systemic Blood Flow, Past, Present, and Future.Pediatric cardiology · 2025Review
- Surgical removal of pulmonary flow restrictors in children with congenital heart disease: What the outcomes reveal.JTCVS open · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeVery high-risk, ductal-dependent or complex two-ventricle patients with associated comorbidities often require pulmonary blood flow restriction as bridge to a more definitive procedure, but current surgical options may not be well-tolerated. An evolving alternative utilizes a fenestrated Micro Vascular Plug (MVP) as a transcatheter, internal pulmonary artery band. In this study, we report a case series and an in-vitro evaluation of the MVP to elicit understanding of the challenges faced with device implantation.
methodsFollowing single-center, retrospective review of eight patients who underwent device placement, an in-vitro flow study was conducted on MVP devices to assess impact of device and fenestration sizing on pulmonary blood flow. A mathematical model was developed to relate migration risk to vessel size. Results of the engineering analysis were compared to the clinical series for validation.
resultsAt median follow-up of 8 months (range 1-15), survival was 63% (5/8), and 6 (75%) patients underwent subsequent target surgical intervention with relatively low mortality (1/6). Occluder-related challenges included migration (63%) and peri-device flow, which were evaluated in-vitro. The device demonstrated durability over normal and supraphysiologic conditions with minimal change in fenestration size. Smaller vessel size significantly increased pressure gradient due to reduced peri-device flow and smaller effective fenestration size.
conclusionDevice oversizing, with appropriate adjustment to fenestration size, may reduce migration risk and provide a clinically appropriate balance between resulting pressure gradient and Qp:Qs. Our results can guide the interventionalist in appropriately selecting the device and fenestrations based on patient-specific anatomy and desired post-implantation flow characteristics.
Indexed as
Identifiers
37580629What OpenQuestion holds
Registered trials