ArticleMaterials (Basel, Switzerland)2023
Biomechanical Investigation of Bone Screw Head Design for Extracting Stripped Screw Heads: Integration of Mechanical Tests and Finite Element Analyses.
Article in Materials (Basel, Switzerland), 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 3 citations in OpenAlex.
- A comparative study of deformation in hexagonal and star-shaped implant screwdriver heads.BMC oral health · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 3 institutions in 3 countries.
Funding
Abstract
Enhancing the design of bone screw head sockets to prevent stripping and improve the torque required for smooth unscrewing is a significant challenge in orthopedic applications. This research aims to establish a quantitative methodology by integrating mechanical testing with finite element (FE) simulations to determine a safe limitation depth for the screwdriver when engaging with the hexagonal socket, thus avoiding stripped screw heads. A FE model was developed to investigate the biomechanical responses of the screw head design. Five custom-made hexagonal sockets were manufactured, and single load torsional tests were conducted to assess the mechanical performance of the screws and drivers. The results from the mechanical tests were compared with the FE simulations, demonstrating a close agreement and confirming the model's validity. Furthermore, additional FE models were created to study the impact of manufacturing tolerances on the socket width and screwdriver width. The findings revealed that the maximum torque to failure for the four designs was lower than the margins specified in ISO 6475. Additionally, increasing the depth of the screwdriver led to higher maximum torque values. This research suggests that the technique of screw insertion, specifically the depth of the driver tool within the screw socket, holds greater importance in preventing stripped screw heads than the design and manufacturing width of the bone screw's hexagonal socket and screwdriver. This confirms the importance of screwdriver engagement inside the bone screw socket to prevent stripped screw heads and sheds light on the added value of maximum torque prediction for future design modifications.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.