A new method to improve RF safety of implantable medical devices using inductive coupling at 3.0 T MRI.
Article in Magma (New York, N.Y.), 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed, 4 citations in OpenAlex.
- Enhancement of in vivo 7T magnetic resonance neuroimaging via flexible metasurfaces.Npj imaging · 2026Article
- Reduction of radiofrequency induced implant heating via flexible metasurface shielding at 7 T.Journal of magnetic resonance (San Diego, Calif. : 1997) · 2025Article
- Hand-held electron spin resonance scanner for subcutaneous oximetry using OxyChip.Magnetic resonance in medicine · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 3 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveTo enhance RF safety when implantable medical devices are located within the body coil but outside the imaging region by using a secondary resonator (SR) to reduce electric fields, the corresponding specific absorption rate (SAR), and temperature change during MRI. MATERIALS AND
methodsThis study was conducted using numerical simulations with an American Society for Testing and Materials (ASTM) phantom and adult human models of Ella and Duke from Virtual Family Models, along with corresponding experimental results of temperature change obtained using the ASTM phantom. The circular SR was designed with an inner diameter of 150 mm and a width of 6 mm. Experimental measurements were carried out using a 3 T Medical Implant Test System (MITS) body coil, electromagnetic (EM) field mapping probes, and an ASTM phantom.
resultsThe magnitudes of B
conclusionA novel method employing the designed SR, which generates opposing magnetic fields to partially shield a sample, has been proposed to mitigate the risk of induced-RF heating at the VoI through numerical simulations and corresponding experiments under various conditions at 3.0 T.