ArticleTechnology and health care : official journal of the European Society for Engineering and Medicine2024
Effect of additional dead space using end-tidal CO2 measurement on ventilating preterm infants: An experimental study.
Article in Technology and health care : official journal of the European Society for Engineering and Medicine, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed, 1 citations in OpenAlex.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundDead space is the part of the airway where no gas exchange takes place. Any increase in dead space volume has a proportional effect on the required tidal volume and thus on the risk of ventilation-induced lung injury. Inserts that increase dead space are therefore not used in small preterm infants. This includes end-tidal CO2 measurement.
objectiveThe aim of this study was to investigate the effect of the end-tidal CO2 measurement adapter on ventilation.
methodsIn an experimental setup, an end-tidal CO2 measurement adapter, three different pneumotachographs (PNT-A, PNT-B, PNT-Neo), and a closed suction adapter were combined in varying set-ups. The time required for CO2 elimination by a CO2-flooded preterm infant test lung was measured.
resultsPNT-A prolonged CO2 elimination time by 0.9 s (+3.3%), Neo-PNT by 3.2 s (+11.6%) and PNT-B by 9.0 s (+32.7%). The end-tidal CO2 measurement adapter prolonged the elimination time by an additional second without the pneumotachograph (+3.6%) and in combination with PNT-A (+3.1%) and PNT-Neo (+3.1%). In conjunction with PNT-B, the end-tidal CO2 measurement adapter reduced the elimination time by 0.3 seconds (-1%). The use of a closed suction adaptor increased the CO2 elimination time by a further second with PNT-Neo (+3.1%) and by an additional two seconds with no flow sensor (+6.9%), with PNT-A (+6.4%) and with PNT-B (+5.5%).
conclusionThe flow sensor had the greatest influence on ventilatory effort, while end-tidal CO2 measurement had only a moderate effect. The increased ventilatory effort levied by the CO2 measurement was dependent on the flow sensor selected. The use of closed suctioning more negatively impacted ventilatory effort than did end-tidal CO2 measurement.