Practical experience commissioning MRI-compatible tandem and ring applicators for use with the Bravos HDR afterloader.

Jessica M Fagerstrom

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

1 author.

Jessica M FagerstromRadiation Oncology, University of Washington, Seattle, Washington, USA.ORCID https://orcid.org/0000-0003-2074-5723

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Five complete MR-conditionally approved ring sets, including fifteen tandems, and two additional rings, were commissioned at an institution intending to use them in an MRI planning environment with a Bravos HDR brachytherapy remote afterloader. Channel length, radiograph, autoradiograph, ring offset, and treatment interrupt measurements were performed, and applicators were assessed in both CT and MRI. During commissioning, one ring was found to be defective and was returned to the manufacturer for a replacement. The eventual complete applicator suite (including the replacement ring) was found to follow the manufacturer-provided specifications, including those delineated in vendor-provided 3D virtual models and those defined within the manufacturer's instructions for use documentation. Based on this work, an offset correction of -0.4 cm will be used for all tested rings using the Bravos system's internal distal dwell position correction feature during treatment preparation. This study reiterated the requirement for careful commissioning of each applicator intended for clinical service considering the intended use and the planned clinical environment and work processes.

Indexed as

BrachytherapyMagnetic Resonance ImagingHumansRadiotherapy DosagebrachytherapycommissioningHDRimage guidanceMRIquality assurancetandem and ring

Identifiers

PMID37469228
PMCPMC10647988

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.