An integrated and fast imaging quality assurance phantom for a 0.35 T magnetic resonance imaging linear accelerator.

James J Sohn, Sara Lim, Indra J Das, Poonam Yadav

Abstract read
In one paragraph

Article in Physics and imaging in radiation oncology, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Evaluation of QUASAR Insight Phantom for daily imaging QA of MRgRT linacs.Journal of applied clinical medical physics · 2026
    Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

James J SohnDepartment of Radiation Oncology, Northwestern Memorial Hospital, Northwestern University Feinberg School of Medicine, IL 60611, USA.
Sara LimDepartment of Radiation Oncology, Northwestern Memorial Hospital, Northwestern University Feinberg School of Medicine, IL 60611, USA.
Indra J DasDepartment of Radiation Oncology, Northwestern Memorial Hospital, Northwestern University Feinberg School of Medicine, IL 60611, USA.
Poonam YadavDepartment of Radiation Oncology, Northwestern Memorial Hospital, Northwestern University Feinberg School of Medicine, IL 60611, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Purpose: Periodic imaging quality assurance (QA) of magnetic resonance imaging linear accelerator (MRL) is critical. The feasibility of a new MRL imaging phantom used for QA in the low field was evaluated with automated image analysis of various parameters for accuracy and reproducibility. Methods and materials: The new MRL imaging phantom was scanned across every 30 degrees of the gantry, having the on/off state of the linac in a low-field MRL system using three magnetic resonance imaging sequences: true fast imaging with steady-state precession (TrueFISP), T1 weighted (T1W), and T2 weighted (T2W). The DICOM files were used to calculate the imaging parameters: geometric distortion, uniformity, resolution, signal-to-noise ratio (SNR), and laser alignment. The point spread function (PSF) and edge spread function (ESF) were also calculated for resolution analysis. Results: The phantom data showed a small standard deviation - and high consistency for each imaging parameter. The highest variability in data was observed with the true fast imaging sequence at the calibration angle, which was expected because of low resolution and short scan time (25 sec). The mean magnitude of the largest distortion measured within 200 mm diameter with TrueFISP was 0.31 ± 0.05 mm. The PSF, ESF, signal uniformity, and SNR measurements remained consistent. Laser alignment traditional offsets and angular deviation remained consistent. Conclusions: The new MRL imaging phantom is reliable, reproducible, time effective, and easy to use for a 0.35 T MRL system. The results promise a more streamlined, time-saving, and error-free QA process for low-field MRL adapted in our clinical setting.

Indexed as

ACR alternativesImagingMagnetic resonance imaging guided radiation therapyMagnetic resonance imaging linear acceleratorMagphanMRIdianQA

Identifiers

PMID37449023
PMCPMC10338140

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.