Evidence map›Paper›PMID 37438580›Full record

ArticleArchives of orthopaedic and trauma surgery2023

In vivo creep and wear performance of vitamin-E-diffused highly crosslinked polyethylene in total hip arthroplasty.

Tsunehiro Matsuoka, Yasuhito Takahashi, Tsunehito Ishida, Toshiyuki Tateiwa, Takaaki Shishido, Kengo Yamamoto

Abstract read
PubMed Publisher
In one paragraph

Article in Archives of orthopaedic and trauma surgery, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Article
  2. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Tsunehiro MatsuokaDepartment of Orthopedic Surgery, Tokyo Medical University, 6-7-1, Nishishinjuku, Shinjuku-ku, Tokyo, 160-0023, Japan.
Yasuhito TakahashiDepartment of Orthopedic Surgery, Tokyo Medical University, 6-7-1, Nishishinjuku, Shinjuku-ku, Tokyo, 160-0023, Japan. yasuhito@tokyo-med.ac.jp.ORCID http://orcid.org/0000-0001-8432-339X
Tsunehito IshidaDepartment of Orthopedic Surgery, Tokyo Medical University, 6-7-1, Nishishinjuku, Shinjuku-ku, Tokyo, 160-0023, Japan.
Toshiyuki TateiwaDepartment of Orthopedic Surgery, Tokyo Medical University, 6-7-1, Nishishinjuku, Shinjuku-ku, Tokyo, 160-0023, Japan.
Takaaki ShishidoDepartment of Orthopedic Surgery, Tokyo Medical University, 6-7-1, Nishishinjuku, Shinjuku-ku, Tokyo, 160-0023, Japan.
Kengo YamamotoDepartment of Orthopedic Surgery, Tokyo Medical University, 6-7-1, Nishishinjuku, Shinjuku-ku, Tokyo, 160-0023, Japan.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionAn acetabular liner thickness of around 6 mm remains the "gold standard" in total hip arthroplasty. Some surgeons have been recommending the use of the thickest possible liner because contact stress and strain in articulating surfaces decrease with increasing the wall thickness. The purpose of this study was to determine whether in vivo creep and wear performance could be enhanced using a thicker liner over the standard thickness in vitamin-E-diffused highly crosslinked polyethylene (HXLPE). MATERIALS AND

methodsOne hundred and twenty-two hips were allocated to age-matched, sex-matched, and body mass index-matched two subgroups implanted either with a 6.8- or 8.9-mm-thick vitamin-E-diffused HXLPE liner against 28-mm cobalt-chrome femoral head, and followed-up for 7 years. Linear and volumetric penetration of femoral head into the liners attributed to creep and wear were analyzed for each group.

resultsCompressive creep strain generated at the initial 6 months was significantly larger in the 6.8-mm group (2.6%) than in the 8.9-mm group (2.2%). The linear steady-state wear observed after 2 years was 0.0019 and 0.0015 mm/year, whereas the volumetric steady-state wear was 0.54 and 0.45 mm

conclusionNo clinical significance for using a thicker liner over the standard thickness (6.8 mm → 8.9 mm) was confirmed in the vitamin-E-diffused HXLPE according to the 7-year follow-up. The wear rates for both thicknesses were very low enough to prevent osteolysis, and no mechanical failure was observed at any follow-up interval. Nevertheless, since the significantly higher strain was seen in the thinner liner, further follow-up is needed to compare the longer term wear and the incidence of osteolysis and component fracture.

Indexed as

Arthroplasty, Replacement, HipHip ProsthesisOsteolysisChildFollow-Up StudiesHumansPolyethyleneProsthesis DesignProsthesis FailureVitamin EVitaminsPolyethyleneVitamin EVitaminsCreepHighly crosslinked polyethyleneTotal hip arthroplastyVitamin EWear

Identifiers

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.